Model 85020A/B
Adjustments
SECTION V
ADJUSTMENTS
5-1.
INTRODUCTION
5-2.
There are no adjustments recommended for regular calibration
or normal use of the 85020A/B Directional Bridge. Hov/ever,
if the
internal bridge microcircuit is replaced, adjustments must be made
to compensate the preamplifier for the individual diode
characteristics. The procedures for replacing the bridge
microcircuit and adjusting the preamplifier are described in
Section VIII, Service, of this manual.
5-1/5-2