Operation
Model 85020A/B
3.
Perform an OPEN/SHORT calibration as follows:
Press 8756A MAIN MENU soft keys
[CAL]
[OPEN/SHORT]
[CHAN 1].
Connect the calibrated short to the test port of the 85020A,
and press
[STORE SHORT] to store the short calibration data.
Remove the short and connect the shielded open to the test
port. On the 8756A press
[STORE OPEN]
to store the open
calibration data. On the 8756A Channel 1 press
[DISPLAY] until
M-MEM is selected. The Channel 1 normalized trace will be
displayed.
4.
Remove the shielded open and connect the 10-dB Attenuator to
the 85020A test port. Leave the other end of the attenuator
unterminated.
5.
On the 8756A press
[CURSOR]. The return loss of the
unterminated pad
(10 dB in each direction) will be displayed.
The display should read -20 dB + 2dB.
Figure 3-2.
85020A Operators Check (2 of 2)
3-6