Technical Specifications Calomat 6E - Siemens ULTRAMAT 6 Series Compact Operating Instructions

Continuous gas analysis
Table of Contents

Advertisement

Technical specifications
7.6 Technical specifications CALOMAT 6E
Parts in gas path wetted by sample gas
Flowmeter
Bypass restrictor (high-pressure version)
Outlet restrictor
Electrical characteristics
Power consumption (unit)
Fuse values
Gas inlet conditions
Permissible sample gas pressure
High-pressure version
– Without built-in pressure regulator
– With built-in pressure regulator
Low-pressure version
Total sample gas flow
High-pressure version
Low-pressure version
Sample gas flow through the O
Sample gas temperature
Sample gas humidity
7.6

Technical specifications CALOMAT 6E

Enclosure
Weight
Dimensions
Electrical characteristics
Power consumption
Fuse values
100 to 120 V
200 to 240 V
Parts in gas path wetted by sample gas
Gas connections
Gaskets (O–rings)
58
Glass
Stainless steel 1.4571
Glass
Approx. 37 VA
100 ... 120 V 1T/250
200 ... 240 V 0.63T/250
 
 
2000 hPa rel.
3000 – 6000 hPa rel.
20 – 30 hPa rel.
 
250 to 500 ml/min (15 to 30 l/h)
125 ml/min (7.5 l/h)
sensor
125 ml/min (7.5 l/h)
2
0 to 50 °C
< 1% RH
Approx. 10 kg
483 mm x 378 mm x 172 mm (W x D x H)
 
Approx. 30 VA
 
1T/250
0.63T/250
Stainless steel 1.4571, pipe diameter 1/4" or 6 mm
FFKM (e.g. Chemraz)
Compact Operating Instructions, 01/2024, A5E45779144002-AB
Series 6 and ULTRAMAT 23