Section VIII
Service
Model 8414A
Table 8-1.
Recommended Test Equipment
Item
Critical Specifications
Recommended HP Model
Oscillator
Frequency
Range: 280 (±50) kHz and 1
(±0.1) kHz
Output Level: variable from 0 to 3.0 Vrms
Output Impedance: 50 to 600 ohms
200CD, 651A
DC Power Supply
Output: -5 Vdc
6213A
Oscilloscope
Vertical:
Minimum bandwidth 5 MHz
Minimum sensitivity 10 mV/cm
Input: dc and ac
Horizontal:
Range: 1 nsec/cm to 5 psec/cm
180A with 1801A and
1821A
Network
Analyzer
No Substitute May be Used.
8410A
with 8411A or
8407A
Sweep Oscillator
Frequency Range: Any frequency within the
operating range of the Network Analyzer.
8690A with RF unit as
required (see Note 1)
20-dB Attenuator
Impedance: 50 ohms nominal
Attenuation: 20 dB ±3 dB
SWR: 1.3 max (1 kHz and 280 kHz)
Connector: BNC
8491A with N to BNC
adapters (see Note 2)
Transducer
Frequency Range: Same as Sweep Oscillator
8740A, 8741A, 8742A,
8743A
or
8745A
(see
Note 3)
Service Cable (sup¬
plied with 8410A or
8407A)
No Substitute May be Used
HP Part No. 08410-6032
High Voltage DCVM
Range: 0 to 6 kV
Accuracy: ±10% of reading
Input Impedance: >12Gohms
410B with 11044A volt¬
age divider probe.
^For 8407A Network Analyzers Model 8601A Sweep Oscillator may be used.
^Oscilloscope 10:1 divider probe, such as HP Model 10003A may be used in place of 20-dB attenuator.
^For 8407A Network Analyzers use HP Model 11561A or BNC Tee.