HP 8414A Operating And Service Manual page 29

Polar display
Table of Contents

Advertisement

Section V
Adjustment Procedures
Model 8414A
Table 5-1.
Recommended Test Equipment
Item
Critical Specifications
Recommended HP Model
Oscillator
Frequency
Range: 280
(±50)
kHz and
1
(±0.1) kHz
Output Level: Variable from 0 to 3.0 Vrms
Output Impedance: 50 to 600 ohms
200 CD, 651A
DC Power Supply
Output: -5 Vdc
6231A
Oscilloscope
Vertical:
(Dual Trace)
Minimum bandwidth 5 MHz
Minimum Sensitivity 10 mV/cm
Input: dc and ac
Horizontal:
Range: 1 psec/cm to 5
n
sec/cm
Network
Analy¬
zer
No substitute may be used
180A
with
1801A
and
1821A
8410A
with
8411A
or
8407A
Transducer
Frequency Range: Same as Sweep Oscillator
8740A,
8741A, 8742A,
8743A
or
8745A
(see
Note 1)
Sweep
Oscillator
Frequency Range: Any frequency within the
8690A with RF unit as
operating range of the Network Analyzer
required (see Note 2)
20-dB Attenuator
Impedance: 50 ohms nominal
Attenuation: 20 dB ±3 dB
SWR: 1.3 max (1 kHz and 280 kHz)
Connector: BNC
8419A with N to BNC
adapters (see Note 3)
Termination
Impedance: 50 ohms ±10% at 278 kHz
Connector: Male BNC
10100A or HP Part No.
1250-0839 with adapter
1250-0831.
^For 8407A Network Analyzers use HP Model 11561A or BNC Tee
^For 8407A Network Analyzers Model 8601A Sweep Oscillator may be used
3
Oscilloscope 10:1 divider probe, such as HP Model 10003A may be used in place of 20 dB attenuator.
5-2