Installation; Power Line Connection - HP 400D Operating And Service Manual

Vacuum tube voltmeter
Table of Contents

Advertisement

T.O. 33A1-12-349-1
Section IL
Paragraphs 2-1 to 2-11
SECTION II
INSTALLATION
2-1.
UNPACKING
AND
INSPECTION.
2-2.
There are no special precautions for unpacking
the voltmeter.
Save the shipping carton and packing
materials for possible storage or reshipment.
When
unpacking, inspect instrument and packing materials
for signs of damage
in shipment.
Make an operation
check as directed in paragraph 2-10 to determine
if
performance is satisfactory.
If there is any indication
of damage, immediately file a claim with the transport
service used or other cognizant authority.
2-3.
LINE VOLTAGE
REQUIREMENT.
2-4.
The voltmeter is wired at the factory for use on
115-volt a-c power.
This voltage may vary +10% without
adverse effect upon voltmeter performance.
The volt-
meter can be wired for use on 230-volt a-c power by
reconnecting the dual primary windings on the power
transformer as shown in the schematic diagram in
Section V.
When using 230-volt power,
change from
a l-amp to a 1/2-amp slow-blow fuse.
If necessary,
provide an adapter for attaching the standard 115-volt
plug on the voltmeter to the 230-volt outlet.
2-5.
POWER LINE CONNECTION.
2-6.
The three-conductor power cable on the voltmeter
is terminated in a polarized three-prong male connector.
The third contact is an offset round pin added to a stand-
ard two-blade connector, which grounds the instrument
chassis when used with the appropriate
receptacle.
To connect this plug in a standard two-contact receptacle,
use
an adapter.
The chassis
ground connection is
brought out of the adapter in a green pigtail lead for
connection to a suitable ground.
2-7.
The power plug normally supplied with the volt-
meter
is made
of molded
rubber and is an integral
part of the power cable.
On certain military contracts,
a modification of the Model 400D, termed the H02-400D,
is equiped with a removable plug having the same pin
configuration but constructed
of corrosion-resistant
material.
In all other respects the H02-400D
is the
same as the Model 400D and carries the same Federal
Stock Number.
00102-2
The lower INPUT and OUTPUT signal terminals
on the panel of the
voltmeter
are
connected
directly to the chassis
of the voltmeter.
Any
voltage
applied
to the lower
terminal
will be
shorted directly to ground.
If the ground con-
nection
in the power
cord is disconnected
by
use of an adapter, the entire voltmeter cabinet
will carry whatever
potential is applied
to the
lower
terminal
and
may
be a hazard
to the
operator.
2-8.
INSTALLATION.
2-9.
The voltmeter is a portable instrument requiring
no permanent installation.
The voltmeter is for bench-
top operation, standing on its rubber feet with its front
panel near the vertical plane.
A bail is provided for
raising the front of the cabinet to obtain a better viewing
angle.
2-10.
OPERATION CHECK.
2-11.
The voltmeter is ready for use as received from
the factory.
The simple check described below can be
made by incoming inspectors to determine if electrical
damage was incurred in shipment.
If more complete
proof of instrument performance is required, the over-all
performance check described in paragraph 5-22 must be
used.
Make a simple performance
check as follows:
a.
Connect
voltmeter
to the power
line through a
variable transformer.
Set transformer for 115 volts,
turn on and allow a five-minute warmup.
b. Measure any sine wave voltage, excepting the power
line, from 0.01 to 300 volts whose exact voltage is known.
Note that the lower INPUT terminal is connected to the
power line ground.
c.
While making the above measurement,
adjust the
line voltage from 103 to 127 volts,
The reading on the
meter must not change by more
than the width of the
pointer.
2-1