Agilent Technologies E8285A Manual page 156

Cdma mobile station test set
Hide thumbs Also See for E8285A:
Table of Contents

Advertisement

Performance Tests
RF Generator Residual FM Performance Test 4
Figure 6-8
Setup for Measurements to 2 GHz Using a Microwave Converter
Audio Analyzer
Measuring Receiver
EXT LO
INPUT
Signal Generator LO
RF
OUTPUT
Procedure
NOTE
Test Sets with firmware revision A.04.5X and above operate only at RF
frequencies from 800 to 1000 MHz and 1700 to 2000 MHz. A text only error
message "Input value out of range" , will be displayed if invalid
frequencies are set.
Steps 1 to 5 in the following procedure apply to both setups (shown in
and
1. On the signal generator (to be used as the measuring receiver's LO):
2. On the measuring receiver:
3. On the audio analyzer:
156
INPUT
MODULATION OUTPUT
INPUT 50Ω
IF OUTPUT
Microwave
Converter
figure 6-8 on page
156).
a. Set the frequency to 11.5 MHz.
b. Set the level to 0 dBm.
a. Reset the instrument.
b. Set the IF to 1.5 MHz (3.2 Special).
c. Set the high-pass filter to 50 Hz.
d. Set the low-pass filter to 15 kHz.
e. Set the measurement mode to FM.
f. If the instrument has an external LO switch, enable the external LO mode
(23.1 Special).
g. If the microwave converter is being used, set the frequency offset mode to
exit the mode (27.0 Special).
a. Reset the instrument.
CALL CONTROL
Answer
Call/
End/
Register
Page
Release
USER
CDMA SCREENS
Cell config
Spectrum
Call
Gen
K1'
control
control
K1
MS report
MS FER
RX test
TX test
K2'
K2
DATA FUNCTIONS
K3'
Ref set
Meter
K3
Increment
Increment
10
set
Assign
Low limit
K4
Release
K5
IBASIC
reset
Shift
Cancel
POWER
RF IN/OUT
DUPLEX OUT
ANTENNA IN
VOLUME
DO NOT apply RF
Max power 2.5W
Max power 60 mW
when in standby
DUPLEX OUT
RF
Signal Generator
INPUT
LO
RF
INPUT
OUTPUT
S:\agilent\e8285\ALR\Book\chapters\perf_test.fm
UTILITIES
Print
Printer
I/O config
Help
Tests
Config
Previous
Message
Preset
Hold
Meas
ANALOG SCREENS
reset
SMS
Address
Authen-
Call
Duplex
Spectrun
Local
tication
control
analyzer
Save
TX
RX test
TX test
AF
Recall
range
analyzer
DATA
Average
Increment
7
8
9
Enter
x 10
High limit
Ghz
4
5
6
dBm
dB
1
2
3
MHz
V
%
0
.
+/-
kHz
mV
s
mW
Yes
No
%
Hz
On/Off
Ratio
dB
µV
W
ms
AUDIO OUT
AUDIO IN
High
Low
Max 12V peak
Max
42V peak
figure 6-7

Advertisement

Table of Contents
loading

Table of Contents