Model 415E
Section П
Paragraphs 2-1 to 2-16
SECTION
Il
PREPARATION
FOR
USE
2-1.
INCOMING
INSPECTION.
2-2,
This instrument
was inspected both mechanically
and electrically before shipment.
To confirmthis,
the
instrument should be inspected for physical damage in
transit.
Also check for suppliedaccessories,
and test
the electrical
performance
of the instrument,
using
the procedure
outlined in Paragraph 5-3.
If there is
damage or deficiency,
see the warranty on the inside
front cover of this manual.
2-3.
INSTALLATION.
2-4.
The Model 415E is fully transistorized; therefore
no special cooling is required.
However,
the instru-
ment should not be operated where the ambient tem-
perature exceeds 55°C (140°F).
2-5.
RACK
MOUNTING.
2-6.
The Model 415E is a submodular unit that when
used alone canbe bench mounted only. However, when
used in combination with other submodular units it can
be benchand/or rack mounted.
The hp combining case
and adapter frame
are designed specifically for this
purpose.
2-7.
COMBINING
CASE.
The combining case
is a
full-module
unit which accepts varying combinations
of submodular units.
Being a full-module unit, it can
be bench ог rack mounted analogous to any full-module
instrument.
An illustration of the combining case is
shown in Figure 2-1.
Instructions for installing the
Model 415E in a combining case are given graphically
in Figure 2-2.
2-8.
ADAPTER
FRAME.
The adapter frameis a rack
frame
that accepts
any combination
of sub-modular
units.
It can be rack mounted only. Ап illustration of
the adapter frame is given in Figure 2-3. Instructions
are given below:
a.
Placethe adapter frame on edge of benchas
shown
іп step 1, Figure 2-4.
b.
Stackthe sub-modularunits inthe frame as shown
in step 2, Figure 2-4.
Place the spacer clamps be-
tween instruments as shown in step 3, Figure 2-4.
c.
Place spacer clamps on thetwo end instruments
(see step 4, Figure 2-4) and push the combination into
the frame.
d. Insert screws oneither side of frame, andtighten
until sub-modular instruments are tight in the frame.
е.
The complete assembly is ready for rack mounting.
2-9. THREE-CONDUCTOR POWER CABLE.
2-10.
To
protect
operating
personnel,
the National
Electrical
Manufacturers'
Association
(NEMA)
rec-
ommends
that
the instrument
panel and cabinet
be
grounded.
А11 Hewlett-Packard
instruments
are
02152-1
equipped with a three-conductor
power
cable which,
when plugged into an appropriate receptacle,
grounds
the instrument.
The
offest pin on the power
cable
three-prong connector is the ground wire.
2-11.
To preserve the protection feature
when орег-
ating the instrument
from a two-contact outlet, use a
three -prong to two-prong adapter and connect the green
pigtail on the adapter to ground.
2-12.
PRIMARY
POWER
REQUIREMENTS.
2-13.
The Model 415E can be operated from anAC or
DC primary power source. The AC source canbe either
115 or 230 volts, 50 to 400 cps.
TheDC sourceis a
24-volt rechargeable battery.
The rechargeable bat-
tery is supplied with option 01 instruments only.
2-14.
For operation from AC primary power, the in-
strument can be easily converted
from
115- to 230-
volt operation.
The LINE
VOLTAGE
switch,
51, а
two-position slide switch located at the rear of the in-
strument,
selects the mode of AC operation.
The line
voltage for which the instrument is set to operate ap-
pears onthe slider of the switch.
А 1/16-ampere,
250
volt fuse is used for both 115- and 230-volt operation.
CAUTION
DO NOT
CHANGE
THE
SETTING
OF THE
LINE
VOLTAGE
SWITCH
WHEN
THE
IN-
STRUMENT IS OPERATING,
2-15.
INITIAL BATTERY
CHECK.
2-16.
The following applies to option 01 instruments
or instruments
that
have
field-installed
batteries.
When the battery is used as the power
source for the
first time,
perform the following steps:
NEC
FILLER
PANEL
|
\
Y
1]
S.
о M
©
——À91
О.
|
о
|
9
u
|
|
4SE-A-IT
Figure 2-3.
Adapter Frame Instrument Combinations
2-1