Table 5-2.
Performance Tests (Cont'd)
5.
BANDWIDTH:
Variable
15 to 130 cps.
Procedure.
Readings
b. Set415Eto LOW, ON, NORM, 0 (RANGE-DB),
with GAIN and VERNIER full clockwise.
"
Act
-
Act
w
44%
Difference
( > 130 cps)
| 2, x
"I
с.
Turn BANDWIDTH
full counterclockwise.
і
d.
Tune Oscillator to peak 415E.
е.
Switch 415Е EXPAND to 0 and retune Oscilla-
tortobe surethat 415E is peakedat center frequency.
|
f.
Adjust GAIN control for a 0 Ф meter reading.
g.
Tune
Oscillator
slightly off tuned
frequency
causing meter reading to drop to exactly 1.5 db and
record Counter reading.
h.
Tune Oscillator back to tuned frequency and
then
off
to other
side of tuned frequency
causing
meter
reading
to drop to exactly
1.5 db.
Record
Counter reading.
і.
The difference
between the readings of steps
g and h should be 15 cps or less.
j. Turn
BANDWIDTH
full clockwise and retune
OSCILLATOR for peak 415E meter reading.
k.
Repeat steps f, є, h above.
m.
The
difference
between
the recorded
fre-
quency
readings,
this time,
should be
130 cps or
greater,
6. SPECIAL 5000 OHM FEED-THRU
TERMINATION
In order to measure the 415E operating noise level, a special load must be used to terminate the INPUT in its
optimum source
impedance.
For the LOW impedance
INPUT,
the hp Model
10100B (100 ohms)should be used.
| For the HIGH impedance INPUT,
a special 5000 ohm termination
must be built as detailedbelow (See Table5-1
for part stock numbers).
|
RESISTOR™
CONTACT
SPRING
|
Procedure
a.
Refer to cut away view to left.
Unscrew
male
BNC and lockwasher
from housing by using a 3/8-
inch open-end wrench and holding housing either in
a vise or with gas pliers.
MALE
BNC
FEMALE
BNC
1250-0095
1250-0251
SOLDER
THESE
HOUSING
If gas pliers are used housing should be pro-
tected with tape or heavy paper.
Ж RESISTOR VALUE:
50,100, OR 5K OHM AS REQUIRED
NOTE:
ENTIRE ASSEMBLY
MINUS RESISTOR IS AVAILABLE
AS 11523-600.
b.
Solder 5000 ohm +10% 1/4 W resistor to BNC.