Guidance and manufacturer's declaration - electromagnetic immunity
The RadiForce series is intended for use in the electromagnetic environment specified below.
The customer or the user of the RadiForce series should assure that it is used in such an environment.
Immunity test
IEC60601 test level
Electrostatic
±6kV contact
discharge (ESD)
±8kV air
EN61000-4-2
Electrical fast
±2kV
transient / burst
for power supply lines
EN61000-4-4
±1kV
for input/output lines
Surge
±1kV line(s) to line(s)
EN61000-4-5
±2kV line(s) to earth
Voltage dips, short
<5% U
interruptions and
for 0.5 cycle
voltage variations on
40% U
power supply input
5 cycles
lines
70% U
EN61000-4-11
25 cycles
<5% U
for 5sec
Power frequency
3A/m
(50/60Hz)
magnetic field
EN61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
6
Compliance level
±6kV contact
±8kV air
±2kV
for power supply lines
±1kV
for input/output lines
±1kV line(s) to line(s)
±2kV line(s) to earth
(>95% dip in U
)
<5% U
T
T
for 0.5 cycle
(60% dip in U
) for
40% U
T
T
5 cycles
(30% dip in U
) for
70% U
T
T
25 cycles
(>95% dip in U
)
<5% U
T
T
for 5sec
3A/m
Electromagnetic environment -
guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that of a
typical commercial or hospital environment.
Mains power quality should be that of a
typical commercial or hospital environment.
(>95% dip in U
)
Mains power quality should be that of a
T
T
typical commercial or hospital environment.
(60% dip in U
) for
If the user of the RadiForce series requires
T
T
continued operation during power mains
(30% dip in U
) for
interruptions, it is recommended that the
T
T
RadiForce series be powered from an
(>95% dip in U
)
uninterruptible power supply or a battery.
T
T
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or hospital
environment.