Rosemount 2051 product certifications

Rev 1.31
7.1 
European directive information
A copy of the EU Declaration of Conformity can be found at the
end of the Quick Start Guide. The most recent revision of the EU
Declaration of Conformity can be found at Emerson.com/Rosemount.
7.2 
Ordinary location certification
As standard, the transmitter has been examined and tested to
determine that the design meets the basic electrical, mechanical,
and fire protection requirements by a Nationally Recognized Test
Laboratory (NRTL) as accredited by the Federal Occupational Safety
and Health Administration (OSHA).
7.3 
Hazardous location certification
NOTICE
Device ambient temperature ratings and electrical parameters may
be limited to the levels dictated by the hazardous location certificate
parameters.
7.4 
North America
E5 USA Explosionproof (XP) and Dust Ignition-proof (DIP)
Certificate
Standards
Markings
Quick Start Guide
2041384
FM 3600: 2022, FM 3615: 2022, FM 3616: 2022,
ANSI/UL 61010-1-2019 Third Edition, ANSI/UL
12.27.01: 2022 (Fourth Edition), ANSI/UL 50E (First
Edition)
XP CL I, DIV I, GP B, C, & D T5;
Seal not required
DIP CL II, DIV I, GP E, F, & G; CL III T5;
T5: (-50 °C ≤ Ta ≤ +85 °C)
Type 4X, IP 68
Optional: single seal
Quick Start Guide