Connecting the FC410

Note
End Of Line (EOL) termination
The FC410 EOL termination DIP switch is default set to EOL active. To change termination
setting see Setting the EOL termination DIP switches (Page 42).
5.4.1

M12 version (for non-Ex areas)

The device is provided with a preformed cable terminated with M12 style stainless steel
weather-proof plugs.
The cable screen is physically and electrically terminated within the body of the plug.
Take care when handling the cable and passing it through cable ducting that the plug is not
subjected to excessive tension (pulling) as the internal connections may be disengaged.
Note
Never pull the cable by the plug - only by the cable itself.
1. Connect the device using the supplied 4-wire cable with M12 connectors.
Note
Grounding
The sensor cable screen is electrically connected to the grounding terminal (PE), only
when the M12 plug is correctly tightened.
Terminal number
1
2
3
4
FC410 with Modbus
Operating Instructions, 02/2016, A5E33120874-AC
Description
24 VDC
0 VDC
B
A
Connecting
5.4 Connecting the FC410
Wire color (Siemens cable)
Orange
Yellow
White
Blue