Fluke 773-II Instruction Sheet page 12

Milliamp process clamp meter
Table of Contents

Advertisement

Sourcing mA
Use mA Source mode whenever it is necessary to source current into a
passive circuit such as a current loop with no loop supply. Source mode
depletes the battery faster than Simulate mode.
To enter Source mode for the 772, see Figure 4:
1.
Insert the test leads into the -mA and +mA jacks.
2.
Press L.
T
3.
Press
To enter mA Source mode for the 773/773-II, see Figure 8:
1.
Insert the test leads into the desired input jacks.
2.
Press L.
T
3.
Press
Simulating mA Output
In Simulate mode, the Meter simulates a current loop transmitter. To
enter Simulate mode, see Figure 9:
1.
Insert the test leads into the +mA and -mA input jacks.
2.
Press L.
T
3.
Press
until Source appears on the display.
until Source appears on the display.
Figure 8. Sourcing mA Output
until Simulate appears on the display.
Fjv10.emf