Exchanging the pilot lamps

-
Pull the mains plug!
-
Remove the screws from the base plate along the case's rim and if necessary
slotted head screws in the feet. Lift the base plate carefully.
-
Pull out the defective lamp.
-
Remove the clamping plugs with silicone hose from the contacts of the
defective lamp.
-
Clip the clamping plugs onto the contacts of the new lamp.
-
Push the new lamp into the housing from outside.
Advices:
-
Please observe the catch to the recess.
-
Avoid contacting of wires and sharp edges (oscillating parts/isolation can be
damaged).
5

Re-starting the unit

After repair, assemble the unit in reverse order as described in item 3.
IMPORTANT
Make sure that all plug connections have been made correctly (for details see wiring plan in app. 3).
Insert the base plate with the generator carefully.
Fix all screws of the unit, fill it with cleaning liquid and switch it on.
If the unit does not work, or if a fault occurs again after only a short period of time, you should return
the unit to the supplier. It is important that you inform him in detail about the circumstances which
lead to the failure, e.g.
-
cleaning liquid used
-
filling-level
-
parameters set - temperature, cleaning duration
-
details about the goods to be cleaned
-
operation time per day or similar