Compensated Correctly
Probe Adjustment Signal
Proper Amplitude of a
1 MHz Test Signal
Once compensated, the probe conveys signals to the input channel accurately, without
artificially attenuating or amplifying frequency components of the signal.
For the most accurate measurements:
Tip
Compensate probes for each channel of the oscilloscope;
Use a compensated probe only with the channel you used to compensate it; and
Compensate your probes frequently.

Programming the PXIe-5163

You can acquire data interactively using InstrumentStudio, or you can use the NI-SCOPE
instrument driver to program your device in the supported ADE of your choice.
16 | ni.com | PXIe-5163 Getting Started Guide
Undercompensated
Probe Adjustment Signal
Reduced Amplitude of a
1 MHz Test Signal
Overcompensated
Probe Adjustment Signal
Increased Amplitude of a
1 MHz Test Signal