917X Series Metrology Wells
Safety Information
DO NOT turn off the instrument at temperatures higher than 100°C. This could create
a hazardous situation. Select a set-point less than 100°C and allow the instrument to
cool before turning it off.
The high temperatures present in Metrology Wells designed for operation at 300°C
and higher may result in fires and severe burns if safety precautions are not observed.
ELECTRICAL HAZARD
These guidelines must be followed to ensure that the safety mechanisms in this instru-
ment will operate properly. This instrument must be plugged into a 115 VAC (230
VAC optional), AC only electric outlet. The power cord of the instrument is equipped
with a three-pronged grounding plug for your protection against electrical shock haz-
ards. It must be plugged directly into a properly grounded three-prong receptacle. The
receptacle must be installed in accordance with local codes and ordinances. Consult a
qualified electrician. DO NOT use an extension cord or adapter plug.
If supplied with user accessible fuses, always replace the fuse with one of the same
rating, voltage, and type.
Always replace the power cord with an approved cord of the correct rating and type.
HIGH VOLTAGE is used in the operation of this equipment. SEVERE INJURY or
DEATH may result if personnel fail to observe safety precautions. Before working
inside the equipment, turn power off and disconnect power cord.
1.3.2

Cautions

To avoid possible damage to the instrument, follow these guidelines:
To properly calibrate the Metrology Well, the "Fan Limit" must be set to "Auto."
DO NOT leave the sleeve(s) in the instrument for prolonged periods. Due to the high
operating temperatures of the instrument, the sleeves should be removed after each use
and buffed with a Scotch-Brite® pad or emery cloth (see the Maintenance section of
the Technical Guide).
Always operate this instrument at room temperature between 5°C and 50°C (41°F
and 122°F). Allow sufficient air circulation by leaving at least 6 inches (15 cm) of
clearance around the instrument. Overhead clearance is required. DO NOT place the
instrument under any structure.
DO NOT interchange inserts between Metrology Well models. The inserts provided
with the instrument as well as any additional inserts ordered are model specific.
Component lifetime can be shortened by continuous high temperature operation.
DO NOT apply any type of voltage to the switch terminals. Applying a voltage to the
terminals may cause damage to the controller.
DO NOT use fluids to clean out the well. Fluids could leak into electronics and dam-
age the instrument.
DO NOT introduce any foreign material into the probe hole of the insert. Fluids, etc.
can leak into the instrument causing damage.