Pressure And Temperature Limits; Operating Voltage Limits - ABB Field 264H Instructions For Installation Manual

Pressure transmitters
Table of Contents

Advertisement

PHASE 1 - PRELIMINARY CHECKS
2.2

PRESSURE AND TEMPERATURE LIMITS

The maximum allowable pressure (PS) and maximum/minimum
allowable temperature (TS) of the pressure transmitter, for
the European market are indicated on the tag plate (ref. D of
Figure 1), in the U.S. please refer to the instrument data sheets
available at: http://138.221.224.36 searching for "264 and
sheet" or from local ABB representatives.
PS and TS are defined according to the European Pressure
Equipment Directive 97/23/EC (PED). For Temperature, the
listing is for the min and max process temperature limits at
which the instrument can be exposed, for Pressure this listing
should be considered the maximum working pressure limit to
which the transmitter can be exposed without damage (to the
transmitter). These limits are valid when the transmitter is
installed according to the instructions in this manual.
WARNING ! - For potentially explosive atmosphere
applications see the temperature range specified in the
certificate/approval relevant to the intended type of
protection.
PS values
models 264GS and 264AS
- 1MPa, 10bar, 145psi for sensor codes C, F
- 0.5MPa, 5bar, 72.5psi for sensor code L
- 6MPa, 60bar, 870psi for sensor code U
- 20MPa, 200bar, 2900psi for sensor code R
- 90MPa, 900bar, 13050psi for sensor code V
- 0.6MPa, 6bar, 87psi for perfluoroelastomer gasket
models 264HS, 264NS, 264HF, 264NF
- 14MPa, 140bar, 2030psi for sensor codes E, G, H, M
- 21MPa, 210bar, 3045psi for sensor codes P, Q, S
- 90MPa, 900bar, 13050psi for sensor code V
Proof pressure
The transmitter can be exposed without leaking to line pressure
of up to:
models 264GS and 264AS
- 1MPa, 10bar, 145psi for sensor codes C, F
- 0.5MPa, 5bar, 72.5psi for sensor code L
- 6MPa, 60bar, 870psi for sensor code U
- 20MPa, 200bar, 2900psi for sensor code R
- 90MPa, 900bar, 13050psi for sensor code V
- 0.6MPa, 6bar, 87psi for perfluoroelastomer gasket
models 264HS, 264NS, 264HF, 264NF
- 28MPa, 280bar, 4060psi for sensor codes E, G, H, M
- 40MPa, 400bar, 5800psi for sensor codes P, Q, S
- 90MPa, 900bar, 13050psi for sensor code V
Sensor code is the 6
character of the product code on
1
th
label Ref. A
Digit 6 at 9
character of the product code on label Ref. A
2
th
Meet ANSI/ISA–S 82.03 hydrostatic test requirements and
SAMA PMC 27.1.
Other 264Hx and 264Nx models using direct mount and/or
remote seal(s) are compliant to SEP (Sound Engineering
Practice).
containment of the process fluid by the transmitter, it is
recommended that any part, related to process fluid
containment, be supplied only by ABB or authorized agents.
The spare parts list is available at: http://138.221.224.36
searching for "SL/26*" or from local ABB representatives.
The storage conditions must be within the following limits,
Lower limit: –50°C (–58°F); –40°C (–40°F) for LCD indicators
Upper limit: +85°C (+185°F) otherwise some material such as
gaskets may be damaged and cannot assure proper fluid
containment.
2.3
The transmitter is designed to provide a current transmission
signal of 4 to 20 mA dc on the same wires that are carrying the
power supply (two-wire transmitter).
The transmitter operates on a minimum voltage of 10.5 Vdc to
a maximum of 42 Vdc and is protected against polarity inversion.
The transmitter operates from 10.5 to 42 Vdc with no load.
The operating voltages range for fieldbus versions, both polarity
independent, are as follows:
- PROFIBUS PA from 9 to 32 Vdc (17.5 Vdc for EEx ia)
- FOUNDATION Fieldbus from 9 to 32 Vdc (for EEx ia
1
applications 24 Vdc max with entity or 17.5 Vdc max with
1
FISCO)
1
Some communication problems may occur, in cases the loop
1
resistance is increased by long wires or additional devices like
1
meters, safety barriers, surge protection, remote indicators,
2
etc.. If this is the case, please refer to the reference manual for
further details (the manual is available at www.abb.com inserting
1
in the "search" tool the keyword "IM/26X") or from local ABB
1
representatives.
1
For EEx ia and intrinsically safe (FM, CSA and SAA) approvals,
the power supply must not exceed 30 Vdc. In some countries
the maximum power supply voltage is limited to a lower value.
The transmitter insulation resistance (terminals to earth) is
>100MΩ at 1000VDC.
1
1
the transmitter as this could break the electrical insulation
1
and damage the electronic part of the transmitter.
1
1
2
1
1
1
WARNING ! - In order to assure the proper

OPERATING VOLTAGE LIMITS