Agilent Technologies Turbo-V 69-8850 User Manual page 287

Remote panel

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 284
Turbo-V Remote Panel User Manual / 87-900-987-01
ANALOG OUT TYPE: permits selection of the type of signal to be
sent to the programmable analogue output. One of the following
signals can be selected:
FREQUENCY: the drive frequency of the pump is shown
POWER: the power supplied to the pump is shown
TEMPERATURE: the temperature of the pump is shown, this
selection is valid only for Tv2k-G and TV3k-T pumps
PRESSURE: the pressure read by the gauge is shown.
START OUT MODE: Selects the operating mode of the START
output. Possible options:
RUNNING: the output is active when the pump is driven
STARTING: the output is active only during acceleration
ramp-up.
SPEED DISPLAY (KRPM/Hz): permits selection of the unit of
measurement of the speed of rotation shown on the display of
the controller. This setting acts only on the measurement shown
on the display. Possible options:
KRPM
Hz
LCD CONTRAST: sets the LCD contrast.
RESET COUNTERS: resets all the counters of the pump (cycles,
time and pump life.
Technical Information
(E)
17
Programming
287/304