Ultra low temperature (ult) freezer
Hide thumbs Also See for DW-86L338J:

Advertisement

Preparation before Use
1. Remove the packaging materials and strings
Remove all packing materials and straps for transportation.
2. Check the supplied accessories
Check the items in the packing box according to the packing list. If they do not match each other,
please contact Haier immediately.
3. Installation environment
There should be at least 30 cm space between the
surrounding walls and the freezer for ventilation.
4. Adjust support legs
Rotate the leveling legs clockwise to extend them
out to support the unit to the fl oor to ensure that
the unit does not move while in usage.
5. Placement
After adjusting and cleaning the unit, do not connect the power cord immediately.
The freezer needs to be placed in its intended location for at least 24 hours before connecting the
power to make sure it will operate normally.
8
Recycling the Rechargeable Battery
The Haier freezer is equipped with a rechargeable battery. This battery is recyclable. When the
battery reaches the end of its life, please contact a local recycling organization for inspection or
properly discard the battery.
Location of the battery
The battery in the electric cabinet is for the power outage alarm.lt is located inside the control box of
the right side of the unit.
There are high voltage components in the control box. To prevent electric shock, only a qualifi ed
technician or engineer can open the cover.
Removal of the battery
1. Turn off the power of the unit and unplug the power cord from the socket.
2. Use a screwdriver to remove the screws on the side panel and take down the side panel.
3. Use a screwdriver to remove the 4 screws on the cover of the control box.
4. Unplug the connecting terminals from the battery.
5. Remove the bracket that fastens the battery Remove the battery.
6. Follow regulations to recycle the battery or discard it properly.
When changing the battery, you must make sure that the brown wire connects to the positive
pole of the battery, and the blue wire connects to the negative pole of the battery. The polarity
must not be reversed. Incorrect polarity can damage the main control board so that it cannot
charge the battery.