Page 2
(since 1972) serving Industries in India & Worldwide thro’ the Manufacture & Supply World-Class Calibration Instruments & Systems like Temperature, Pressure & Signal Calibrators, Black Body Calibration Sources, Pneumatic & Hydraulic Hand Pumps, Dead Weight & Comparison Testers, Calibration Test Benches, etc. NAG/IM/PHP 10-40/01/22...
Page 3
Dear User, Thank you for selecting Nagman’s Pneumatic Hand Pump and becoming a proud owner of this Calibration Instrument. We have strived hard to ensure the accuracy of the contents of this manual. We would appreciate any suggestions/feedback to correct any errors noticed and to improve...
INTRODUCTION Nagman’s calibration hand pump is designed to manually generate pressure between -0.90 upto 10/25/35/40 bar (depending upon models) of vacuum and pressure for quick and accurate calibration of pressure gauges and other pressure measuring Instruments. This pump is manufactured with high quality components for calibrating Analog / Digital gauges by comparison method.
Upto 1 mbar Gauge / Indicator 3/8” (F) BSP Connection Port Test Connection Port 1/4" (F) BSP Stabilization time 1-2 minutes Instrument Dimension 230 x 140 x 65 mm (L x B x H) Instrument Weight 750 gms. (approx.) NAG/IM/PHP 10-40/01/22...
Knurling Thumb Nut (2) (Over Pressure Protection): To adjust maximum output pressure of the pump turn the knurling thumb nuts (2) to increase or decrease the stroke length so that the maximum pressure is controlled per stroke. NAG/IM/PHP 10-40/01/22...
7. SAFETY INSTRUCTIONS S. No. Symbol Description Read the user manual before operating the instrument. Warning - conditions that may pose hazards to the user. Caution-conditions that may damage the instrument. Special Information NAG/IM/PHP 10-40/01/22...
After calibration is completed, open the Release valve gradually in anticlockwise direction to remove pressure / vacuum before disassembling the measurement system (Master / Test Instrument). Remove the master and the test instrument from the pump. NAG/IM/PHP 10-40/01/22...
Chances are there that one of the seals in the PHP is leaking and needs to be replaced, if the applied pressure still cannot be obtained. If functionality of Hand Pump is not normal, please do not Continue Pumping. NAG/IM/PHP 10-40/01/22...
Page 15
Parts are replaced / repaired using spare parts which are not identical to those recommended by the manufacturer. Nagman‘s liability is restricted to errors that originated from the factory. For more details, write to : NAGMAN INSTRUMENTS AND ELECTRONICS...