ABB 8242 Instruction Manual page 43

Colorimetric phosphate monitor
Hide thumbs Also See for 8242:
Table of Contents

Advertisement

SECONDARY CALIBRATION SEQUENCE PAGE 2.5
Time to compensation = xx min
Abort SEC calibrations = NO
CALIBRATION RECOVERY SEQUENCE
Reading =
Time to end of sequence =
Zero offset = xx.x <units>
Calibration factor = x.xx
A3 Page 3 - Set Up Instrument
SET UP INSTRUMENT
Enter security code
SET UP INSTRUMENT
Software issue =
Control temperature =
Display units =
Display units as =
Alter setup security code =
CLOCK SETUP.
Date (DD/MM/YY) =
Time (HH:MM)
WARNING: Adjusting any of the above
time parameters will alter the next
Auto Calibration Date.
SET UP MULTISTREAM SYSTEM
Stream in use
= 1
ON
Stream sequence= xxxxxxxxxxxx
A4 Page 4 - Set Up Output Currents
SET UP CURRENT OUTPUT
Stream 1 O/P range = 0 to xxx<units>
Stream 2 O/P range = 0 to xxx<units>
Stream 3 O/P range = 0 to xxx<units>
Stream 4 O/P range = 0 to xxx<units>
Stream 5 O/P range = 0 to xxx<units>
Stream 6 O/P range = 0 to xxx<units>
Reading = xxx<unit>
PAGE 2.6
xxx <units>
xx min
PAGE 3.0
xxxx
PAGE 3.1
x
xx.x'C
<units>
xx
xxxx
PAGE 3.2
xx/xx/xx
=
xx:xx
PAGE 3.3
2
3
4
5
6
ON
ON
ON
ON
ON
PAGE 4.0
APPENDIX A – MULTI-STREAM PROGRAMMING...
Reading during calibration prior to compensation.
Remaining time to the end of the Secondary Calibration
sequence.
When set to YES the sequence is aborted. The original
calibration factor value is maintained.
Reading during recovery back onto the sample. The display
shows the reading which incorporates new calibration values.
Remaining time to end of sequence.
See Programming Page 1.1.
Enter the value of the previously entered security code.
Shows the current software issue level.
Set the required control temperature within the range 20 to
45C in 0.1C increments. This temperature should be set to
or 5C above the maximum ambient temperature expected.
Set the required display units for phosphate concentration
(ppm, mg/l or mg/kg).
Set the display type as P (Phosphorus) or
(Orthophosphate).
Enter setup security code (up to four digits) if required.
Set the current (real) date and time respectively.
Enables streams to be selected/de-selected.
Used to set the stream sampling order.
Note. If one stream only is selected the monitor behaves
as a single stream version.
Set the current output to any range between the following
maximum and minimum limits:
3–
PO
4
0 to 3
and
0 to 60
3 –
PO
4
P
0 to 1
and
0 to 20
41