Triggering - The Trigger Subsystem - Agilent Technologies 8163A Programming Manual

Lightwave multimeter lightwave measurement system lightwave multichannel system
Hide thumbs Also See for 8163A:
Table of Contents

Advertisement

Triggering - The TRIGger Subsystem

Hardware
Trigger Rearming
Triggering
trig:inp
trig:inp:rearm
IGNore
-
SMEasure
ON
CMEasure
ON
SMEasure
OFF
CMEasure
OFF
Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement System, & Agilent 8166A Lightwave Multichannel
System, Third Edition
Triggering - The TRIGger
Subsystem
The TRIGger Subsystem allows you to configure how the instrument
reacts to incoming or outgoing triggers.
Table 7
Triggering and Power Measurements
Software Triggering
init:imm
init:cont
One power
Automatically performs power
measurement is
measurements.
performed.
Every hardware trigger starts a new power mea-
surement.
The first hardware trigger starts a new power mea-
surement. Further hardware triggers are ignored
until you send trig:inp:rearm again.
Measurement Operations & Settings
Data Acquisition Functions
sens:func:stat
MINMax
LOGGing|STABility
Automatically performs power
measurements until the func-
tion is finished.
Every hardware trigger starts
a new power measurement
until the function is finished.
The first hardware trigger
starts the function. Subse-
quent power measurements
are automatically performed
until the function is finished.
Every hardware trigger starts
a new power measurement
until the function is finished.
The first hardware trigger
starts the function. Subse-
quent power measurements
are automatically performed
until the function is finished.