6. For simulated distillation or other special software:
– press
– Press
If the analyzer icon is still named "Analyzer", right click on it to rename it. The foregoing
procedure must only be done once. Double click on the icon to launch the analyzer in
EZChrom. For details, see the EZChrom release notes.
7. Import the EZChrom method:
press the
there.
4.11

Run Chromatograms and Make Method Adjustments if Necessary

If any changes were made to EPC setpoints in the preceding steps, update the setpoints in the
method and export the edited method back to the Maxum II.
See the user example
Chromatograph Portal Reference Manual .
4.12

Calibrate the Maxum II

Follow the calibration instructions shipped with the analyzer.
More information on various aspects of calibration are given in the Maxum II Gas
Chromatograph Portal Reference Manual
109762016/)
The following excerpt from that manual,
example:
Overview:
Calibration is another change to the method similar in many ways to the changes described in
the previous example for changing a method. Calibration is accessed from the Method Peaks
screen described in the previous section. Calibration functions are accessed on this screen by
clicking the box labeled "Calibration".
Installation Manual
Installation Manual, 12/2019, 2000595 Rev 1
Analysis Options > [check the appropriate box] > OK .
Auto-Configuration > OK > OK .
File > Method > Import . Highlight the desired method and
Import button. Create a folder on the hard drive and save a copy of the method
Changing a Method in the How To section of the Maxum II Gas
(https://support.industry.siemens.com/cs/document/
Calibrating a Method in the How To section, is an
Initial Startup Procedures
4.12 Calibrate the Maxum II