5.
Cut the GC liner per the template in Figure 2, and then secure the
mounting plate onto the oven with the screws provided.
6.
Remove the right side panel of the GC to expose the PCB mounted
connector receptacles. Attach the Burner mount's power connector to the
GC's temperature control circuit. Consult your GC's operation or service
manual to confirm proper placement.
7.
Position the Dual Plasma Burner (column end down) into the tapered
fitting of the heated base, with the lower hydrogen line and pin aligned
with the slot in the heated base. The Burner should be secure when
properly positioned.
8.
Follow the instructions in the "Installation" chapter of the Operation and
Maintenance Manual for connecting the Burner to the Dual Plasma
Controller.