Accessory G6600-60621
Purpose
To install the Dual Plasma Burner mounting kit on a gas chromatograph (GC)
unit.
Scope
This procedure describes how to install the Dual Plasma Burner mounting kit.
Safety
Make sure all devices are powered off while performing the installation.
Materials
A Dual Plasma Burner mounting kit
The Dual Plasma Burner
A Philips screwdriver
3 mm Allen wrench
Sheet metal cutting tool(s)
Procedure
1.
Remove the mounting kit from the packaging.
2.
Make sure the gas chromatograph (GC) is powered off and gas supplies
are off.
3.
On the GC, remove the cover from the detector area to expose the top of
the GC, through which a detector is normally mounted. Find the
appropriate mounting location based on Figure 1.
4.
Align the Burner mounting plate with the mounting screw holes on the
GC. Clear the hole into the GC oven of interfering insulation (approx
1.8 inch diameter).

Advertisement

loading

Summary of Contents for Agilent Technologies G6600

  • Page 1 Installing the Dual Plasma Burner Mounting Kit on a Shimadzu 2014 GC Accessory G6600-60621 Purpose To install the Dual Plasma Burner mounting kit on a gas chromatograph (GC) unit. Scope This procedure describes how to install the Dual Plasma Burner mounting kit.
  • Page 2 Cut the GC liner per the template in Figure 2, and then secure the mounting plate onto the oven with the screws provided. Remove the right side panel of the GC to expose the PCB mounted connector receptacles. Attach the Burner mount’s power connector to the GC’s temperature control circuit.
  • Page 4 Copyright © 2007 Agilent Technologies, Inc. 2850 Centerville Road Wilmington, DE 19808-1610 Printed in the USA Feb 2007...

This manual is also suitable for:

60621