Wahl Digi-Stem DST600 Series User Manual

Hide thumbs Also See for Digi-Stem DST600 Series:

Advertisement

Quick Links

DST600 Series
Digi-Stem
Thermometers
®
User Manual
04/12/12
Wahl Instruments, Inc.
WD1032 Rev L (05/24/16)
234 Old Weaverville Road
Asheville, NC 28804
For use with models:
Toll Free 800-421-2853
DST600
Phone 828-658-3131
DST610, DST611, DST620 & DST621
Fax 828-658-0728
DST640
www.palmerwahl.com
DST650, DST651, DST660, DST661
Including "I", "R" and "P" versions

Advertisement

Table of Contents
loading

Summary of Contents for Wahl Digi-Stem DST600 Series

  • Page 1 DST600 Series Digi-Stem Thermometers ® User Manual 04/12/12 Wahl Instruments, Inc. WD1032 Rev L (05/24/16) 234 Old Weaverville Road Asheville, NC 28804 For use with models: Toll Free 800-421-2853 DST600 Phone 828-658-3131 DST610, DST611, DST620 & DST621 Fax 828-658-0728 DST640 www.palmerwahl.com...
  • Page 2: Table Of Contents

    Or email: register@palmerwahl.com 1. Applications, Key Features, and Models The Wahl DST6xx series is designed for temperature monitoring applications, specifically in the food, dairy and pharmaceutical markets, however it is well suited for any application requiring a high degree of accuracy and reliability.
  • Page 3 DST600 Series USER MANUAL Self-checking Electronic Circuitry: Meter repetitively checks the electronics • accuracy by measurement of on-board ultra-stable, low ppm precision resistors against embedded firmware values and displays “err” in the event the electronics is out by more than ±0.5°F (0.3°C). The self-checking electronic circuit may be tested by rapidly changing the temperature of the sensor, such as placing from room temperature into ice bath or heated process.
  • Page 4: Ce Approval

    3. Installation Caution! See important information regarding Lithium Metal Batteries in Wahl document #WD1053, before proceeding! Your unit was shipped partially assembled. Installation of the battery by the end- user is necessary as transportation regulations prohibit shipping units with the battery installed.
  • Page 5 H-frame to seal the panel opening. 3.3.2. For detailed installation instructions see Wahl document WD1100 and reference Wahl drawing 13615, both included in section 11 of this manual. 3.4. Wire Connections – In the event that your probe was shipped disconnected from the meter, the following describes the probe wiring and installation process.
  • Page 6 4 to 5 in-lbs. of torque. 3.5. Ferrite Installation The Wahl DST600 CE compliance requires installation of 2 clamp-on ferrites around the internal wiring. For other DST6xx series, 2 clamp-on ferrites are included to minimize RFI susceptibility. These ferrites should be placed around the internal wiring.
  • Page 7: Battery Installation / Replacement / Battery Life

    4. Battery Installation / Replacement / Battery Life The DST6xx Series use a single 3.6V C-cell Lithium Thionyl Chloride battery, Wahl Catalog # DSA3062. The DST6xxI uses a single 3.6V AA-cell Lithium Thionyl Chloride battery, Wahl part number 12234-03. Low battery is indicated by “LOW BATT”...
  • Page 8: Operation

    DST600 Series USER MANUAL 4.1.2. Remove old battery and dispose of in accordance with local, state and federal regulations. 4.1.3. Insert new battery with polarity as indicated on battery holder. Positive terminal should be at the top of the PCB. 4.1.4.
  • Page 9 DST600 Series USER MANUAL 5.4. Error Codes - During normal operation, the DST6xx continually performs diagnostic testing on the sensor lines. Errors are indicated by the following error codes: Error Code Description Reading is above meters usable range Reading is below meters usable range or sensor is shorted CbL1 Indicates cable 1 is open (J1 pin 1) CbL2...
  • Page 10: Calibration

    Quality Control System requires either “no positive error” or a lesser amount of positive error, for example, +0.2°F maximum error. It is possible on the Wahl DST6xx series thermometer to offset the error to achieve this accuracy. It should be noted that in doing this, there may be a slightly larger negative error at some temperatures;...
  • Page 11 DST600 Series USER MANUAL To achieve this we will be using the programmable RO value in the DST6xx Series TID, which is accessed using the DSTCAL program. The R0 value (R0 stands for resistance at 0°C), is a variable in the equation for converting resistance to temperature for platinum RTD’s and is factory set to match the specific value of the probe connected to the DST6xx.
  • Page 12: Specifications - Dst600 / Dst640 Series

    DST600 Series USER MANUAL Wahl ® DST600 Meter Specifications and 7. Specifications – DST600 / DST640 Series Accessories DST600 / DST640 S RTD M ERIES ETER PECIFICATIONS DST600 DST600R DST600I Standard Series All Stainless Steel, Upgraded DST640 DST640R DST640I Board, RFI Protection Series...
  • Page 13: Dst610/650, Dst611/651, Dst620/660, And Dst621/661

    DST600 Series USER MANUAL 8. DST610/650, DST611/651, DST620/660, and DST621/661 Models DST610/650, DST611/651, DST620/660 and DST621/661 are variations of the DST600/640 that include an isolated secondary sensor housed in the same probe shank. This enables the user to have an independent signal from the same probe for use by a controller or recorder.
  • Page 14 This option requires the use of a connecting cable or connector (ordered separately). Connector wiring are shown in Figure A2. 8.2.1. Optional IP68 rated field wireable connectors and cables: Wahl P/N Description 12075-07 Straight field wireable connector, 4-6 mm cable gland...
  • Page 15 DST600 Series USER MANUAL 8.3. DST620/660 – Consists of DST600/640 meter with an isolated secondary sensor and 4-20 mA programmable transmitter. Transmitters are programmed at the factory. The DST620/660 provides an isolated current output from the same probe shank. This reduces the number of penetrations made into the process and reduces error due to non- uniformity of the process.
  • Page 16 DST600 Series USER MANUAL 8.4.1. Optional IP68 rated field wireable connectors and cables: Wahl P/N Description 12075-07 straight field wireable connector, 4-6 mm cable gland 12075-08 straight field wireable connector, 6-8 mm cable gland 12075-09 right angle field wireable connector, 4-6 mm cable gland...
  • Page 17: Specifications - Dst650, Dst651, Dst660, Dst661

    DST600 Series USER MANUAL 9. Specifications – DST610/650, DST611/651, DST620/660, DST621/661 DST610 - 620 S / DST650 - 660 S RTD M ERIES ERIES ETER PECIFICATIONS DST610 DST611 DST620 DST621 Standard Series All Stainless Steel Series, Increased DST650 DST651 DST660 DST661 RFI Protection Model...
  • Page 18: Service

    11. Panel Mount Instructions The Panel Mount system for Wahl Digi-Stem is designed for use with panels from 1/8” to 3/8” thickness. The system uses a square O-ring to seal around the meters H-Frame and the panel. The unit is secured by a U-bracket with 2 #8 wing nuts and lock washers.
  • Page 19: 516 Panel Mount Drawing

    DST600 Series USER MANUAL 12. 13516 Panel Mount Drawing 19 of 20...
  • Page 20: See Addendum - Wd1053 Battery Install & Shipping Instructions

    DST600 Series USER MANUAL 13. See Addendum – WD1053 Battery Install & Shipping Instructions ISO 9001:2008 CERTIFIED QUALITY MANAGEMENT SYSTEM 234 Old Weaverville Road, Asheville, NC 28804 Toll Free: 800-421-2853 • Phone: 828-658-3131 • Fax: 828-658-0728 www.palmerwahl.com info@palmerwahl.com 20 of 20...

Table of Contents