Pentax OS-H4 Assembly Instructions

Pentax OS-H4 Assembly Instructions

Water bottle

Advertisement

Quick Links

WATER BOTTLE ASSEMBLY OS-H4
1. INTENDED USE
The PENTAX water bottle assembly, OS-H4 is- intended to be used with compatible PENTAX flexible endoscopes and
PENTAX air sources (video processor, light source, etc.) for delivery of air and/or water during endoscopic procedures.
NEVER use the water bottle for any purpose other than that for which it has been designed.
Since endoscopic accessories, including the water bottle are designed to be used in conjunction with other medical devices,
the effectiveness of an accessory especially a reusable one, is dependent upon a number of factors, including the condition of
the endoscope and the care and maintenance of the accessory and related devices.
WARNING:
The water bottle assembly identified in this manual is a reusable medical device. Since it is packaged nonsterile, it must
be sterilized BEFORE initial patient use. Subsequently, it should be subjected to an appropriate cleaning and sterilization
process at least on a daily basis or more frequently depending upon the patient and/or type of endoscopic procedure.
2. NOMENCLATURE
NOTE:
If the water bottle assembly has been handled roughly, the water feeding tube inside the hose may be damaged or
disconnected at the A/W connector for the endoscope. To test, remove the cap and using a syringe, inject water into the
water feeding stem. If the water comes out of both the center hole of the A/W connector and the series of holes around
the center hole, the water feeding tube is disconnected. Use another water bottle assembly.
3. DIRECTIONS FOR USE
1) Fill the bottle approximately 2/3 full with sterile water.
2) Screw the bottle cap snugly onto the bottle.
NOTE:
Do NOT use the new PENTAX OS-H4 water bottle cap with the older OS-H2 water container/bottle. Although the cap
may appear to fit onto the bottle, air may escape resulting in insufficient pressure and flow of air and water during the
endoscopic procedure. Both the PENTAX water bottle cap and bottle (container) are identified by their appropriate
model designation. Ensure that an OS-H4 cap is used only with the OS-H4 water container/bottle.
Do NOT over-tighten the bottle cap. Over-tightening can cause the bottle cap to break.
3) Set the Air/Water - Drain lever to A/W position.
4) Insert the air pipe stem into the processor or light source and press until the water bottle 'clicks' into position.
A/W
DRAIN
Air/Water-
Drain Lever
Bottle
- 1 -
A/W Hose
A/W Connector
Holder
Air Pipe Stem
Water Bottle Cap
Water Feeding Stem

Advertisement

loading

Summary of Contents for Pentax OS-H4

  • Page 1 NOTE: Do NOT use the new PENTAX OS-H4 water bottle cap with the older OS-H2 water container/bottle. Although the cap may appear to fit onto the bottle, air may escape resulting in insufficient pressure and flow of air and water during the endoscopic procedure.
  • Page 2 The sterilization parameters are only valid with sterilization equipment that is properly maintained and calibrated. Use appropriate heat process indicators and/or biological monitors as recommended by the manufacturer of the sterilizer. For more detail, refer to the manual supplied with the PENTAX endoscope. - 2 -...
  • Page 3 Steam Sterilization (Autoclaving) PENTAX water bottle assemblies, Model OS-H4 have been designed to withstand high-pressure steam sterilization procedures providing the following parameters are adhered to: Sterilizer Type: Prevacuum Temperature: 132 ~ 135 °C (270 ~ 275 °F) Time: 5 minutes During steam sterilization, ensure that the cap and tubing section have been removed from the water bottle container.
  • Page 4 Explanations of the symbols used in the 2-7-5 Naka-Ochiai, Shinjuku-ku, operating manual, on the devices or on the Tokyo 161-8525, Japan packagings. HOYA Corporation PENTAX Life Care Tokyo Office 3-16-3 Higashi-Ikebukuro, Toshima-ku, Symbol for "MANUFACTURER" Tokyo 170-0013, Japan Tel.: ++81-3-5953 2592...