12 Commissioning

12 Commissioning

12.1 Initialization of 1272, 1273

The instrument sensors must be calibrated prior to commis-
sioning with the adjustment potentiometers. The adjustment
must be carried out with the signal receiver at 20% (= 7.2 mA)
and 80% (= 16.8 mA) volumetric flow, in order to compensate
for the difference between the linear output signal and the
non-linear scaling of the flowmeters.
3
2
1
1. Connect mA-measurement device as specified on the in-
strument sensor.
2. Allow 20% volumetric flow to run through the flowmeter.
3. Press "min" key 1.
ð LED display 2 flashes green, the value has been accep-
ted.
ð LED display 2 goes off.
4. Allow 80% volumetric flow to run through the flowmeter.
5. Press "max" key 3.
ð LED display 2 flashes green, the value has been accep-
ted.
ð LED display 2 lights up green, the product is initialized and
ready for operation.

12.1.1 Resetting the product

3
1
To carry out a reset of the product, press the 1 "min" and 3
"max" keys simultaneously.
The reset is completed when the status LED lights up red.
GEMÜ 127x
24 V
14 / 20
A reset must be performed in the following scenarios and
then the product must be reinitialized:
If a user error has occurred when initializing the product.
If the initialization result is not satisfactory.
If another float/variable area flowmeter is used.

12.2 Commissioning the flowmeter

● Commission the flowmeter (see installation, operating and
maintenance instructions for the flowmeter).