Kenwood GP-610A Instruction Manual page 6

Gp-ib adapter
Table of Contents

Advertisement

Temperature c o e f f i c i e n t
Rising time
(10 to 90%, 10 kiloohm load)
[ D i g i t a l I/O]
OUTPUT ON/OFF signal output
Interrupt signal (CV/CC) signal input
Interrupt signal (POWER OFF) signal input : TTL level High: OFF
[Operating condition]
Specified temperature/humidity range: 0 to 40 °C, 80% or less
[ I n s u l a t i o n d i e l e c t r i c strength]
Between GP-1B connector outputs
(Analog output A, B, C)
Between analog output
Between the output and case
Between the input power supply and case
Between the input power supply and output : Same as above
[ I n s u l a t i o n r e s i s t a n c e ]
Between the input power supply and case : 30 megohms or more at 500 VDC
Between the output and case
[Power supply]
100 VAC +/- 10%, 50/60 Hz. S w i t c h i n g to 120 V, 220 V, or 240 V (maximum
250 V) can be done.
[Power consumption]
15 W, 19 VA ( a t 100 V r a t i n g )
: 50 ppm/°C ( t y p i c a l )
.
i n n
1
1 U U / / S
r l e S S
°
: TTL level High: ON
: TTL level High: CV
: 500 VAC for 1 minute
: Same as above
: Same as above
: 1 kVAC for 1 minute
: same as above
- 3 -
Low : OFF
Low : CC
Low : ON