• Ful1-sea 1e adjustment
The f u l l - s c a l e voltage ( c u r r e n t ) output corresponding to the controlled power
supply i s set from the GP-IB bus.
FS semi-solid v a r i a b l e r e s i s t o r on the panel of GP-610A/I).
the COARSE (coarse adjustment) and FINE ( f i n e adjustment) VRs are provided for
f u l l - s c a l e adjustment.
When channel A i s used, use VR ( 8 ) and ( 9 ) on the panel.
When channel B i s used, use VR (11) and (12) on the panel.
When channel C i s used, use VR (14) on the panel.
Note: Models whose rated f u l l - s c a l e voltage i s 36 V or more i n d i c a t e s the
lowest d i g i t i n u n i t s of 100 mV in f u l l - s c a l e ind icat ion.
when a v o l t a g e which i s 20 V or l e s s i s s e t and the lowest d i g i t i s
indicated in u n i t s of 10 mV a f t e r f u l l - s c a l e adjustment, the voltage set
from the c o n t r o l l e r may d i f f e r by s e v e r a l d i g i t s from power supply
indication on these models.
Note: When the PA s e r i e s i s c o n t r o l l e d by t h i s component with OP-13 incor-
porated, set the voltage/current sett ing v a r i a b l e r e s i s t o r on the panel
of the PA s e r i e s power supply to maximum.
Perform offset adjustment and f u l l - s c a l e adjustment in this s t a t e .
In t h i s case, i f the s e t t i n g v a r i a b l e r e s i s t o r on the panel of the PA
power supply i s operated during control through the GP-IB bus, the out-
put s e t t i n g value changes.
Power
output
CP-IB input
F u l l - s c a l e adjustment i s done by using the
Therefore, care must be taken.
- 2 9 -
For output A and B,
Therefore,