Contents General Information Instrument Identification 5975 MSD versions Important Safety Warnings Many internal parts of the MSD carry dangerous voltages Electrostatic discharge is a threat to MSD electronics Many parts are dangerously hot Hydrogen Safety Chemical Safety Safety and Regulatory Certifications...
Page 4
Voltage ranges of major components Power configurations Power requirements Power plugs and cord Other electrical considerations Air Conditioning and Environmental Requirements Temperature, humidity, and altitude Airborne dust Exhaust venting Fume (exhaust) hood Carrier and Reagent Gas Requirements Regulators, tubing, and fittings Laboratory Supply Requirements Cleaning solvents Data system supplies...
Page 5
5975 Series Mass Selective Detectors Site Preparation Guide General Information Instrument Identification 6 Important Safety Warnings 7 Hydrogen Safety 9 Chemical Safety 10 Safety and Regulatory Certifications 11 Cleaning 14 Recycling the Product 14 This section provides information on how to identify specifically what type of MSD you are working with as well as how to properly clean and dispose of the instrument.
10-character serial numbers. Write the serial number of your 5975 MSD here for reference: 5975 MSD versions Table 1 shows the kinds of 5975 MSDs available. The product number on the serial number label identifies your MSD. Table 1...
Important Safety Warnings Before moving on, there are several important safety notices that you should always keep in mind when using the 5975 Series Mass Selective Detector. Many internal parts of the MSD carry dangerous voltages If the MSD is connected to a power source, even if the power switch is off, potentially dangerous voltages exist on: •...
General Information The use of incorrect or makeshift fuses or the short-circuiting of fuse holders creates a shock hazard for the operator and can damage the instrument. Replace fuses only with fuses of identical current rating and type. Excessive fluctuations in the line voltage can create a shock hazard and can damage the instrument.
When mixed with air, hydrogen can form explosive mixtures. A detailed discussion of hydrogen safety appears in Chapter 1 of the Agilent 5975 MSD Troubleshooting and Maintenance Manual. Read this material before using hydrogen with the MSD. When using hydrogen (H WA RN ING gas can flow into the oven and create an explosion hazard.
General Information Chemical Safety The foreline pump exhaust and split vent exhaust will contain traces of the WA RN ING chemicals you are analyzing. These could potentially be toxic. Vent the foreline pump exhaust and split vent exhaust outside your laboratory or into a fume hood. Be sure to comply with all local environmental regulations.
• AUS/NZ This ISM device complies with Canadian ICES-001. Cet appareil ISM est conforme a la norme NMB—001 du Canada. The 5975 Series MSD is designed and manufactured under a quality system registered to ISO 9001. Information The Agilent Technologies 5975 Series MSD meets the following IEC (International Electro-technical Commission) classifications: Equipment Class I, Laboratory Equipment, Installation Category II, Pollution Degree 2.
Failure to comply with these precautions violates safety standards of design and the intended use of the instrument. Agilent Technologies assumes no liability for the customer’s failure to comply with these requirements.
4 Make sure that all peripheral devices are also certified. 5 Make sure that appropriate cables are used to connect the device to peripheral equipment. 6 Consult your equipment dealer, Agilent Technologies, or an experienced technician for assistance. 7 Changes or modifications not expressly approved by Agilent Technologies could void the user’s authority to operate the equipment.
General Information Cleaning To clean the unit, disconnect the power and wipe down with a damp, lint-free cloth. Recycling the Product For recycling, contact your local Agilent sales office. Site Preparation Guide...
5975 Series Mass Selective Detectors Site Preparation Guide Site Preparation Overview 16 Space and Weight Requirements 20 Communications Requirements 22 Electrical Requirements 23 Air Conditioning and Environmental Requirements 28 Carrier and Reagent Gas Requirements 31 Laboratory Supply Requirements 33 Receiving the System 37 Installation and Verification 39 This section describes how to properly prepare your site for a new MSD.
Site Preparation Overview Before the 5975 Series MSD system can be installed, the site must be properly prepared. Site preparation includes, but is not limited to, ensuring that adequate facilities are available. Among the site requirements are: • Adequate space is available for the MSD system.
Agilent Technologies representative Agilent responsibility An Agilent Technologies service representative will install the MSD and verify its performance. The service representative’s responsibilities are limited to: • Unpacking the MSD system and verifying that all components are present and undamaged •...
Page 18
• Installing helium carrier gas or reagent gas tanks and regulators • Connecting or verifying the performance of hardware and software not provided by Agilent Technologies Items that do not include installation by Agilent Technologies must be installed by the NO TE customer.
Site Preparation Other documentation Additional information is contained in the following documentation: • 5975 Series MSD Troubleshooting and Maintenance Manual • 5975 Series MSD Operation manual • 6890 Series GC manuals • GC accessories (autosampler, etc.) manuals • G1701EA MSD Productivity ChemStation software manuals and online help •...
The hose is stiff and cannot be bent sharply. Do not put the foreline pump on your laboratory bench if vibration-sensitive equipment CAU TI O N is located on the bench. 5975 Series MSD ALS tray Figure 1 Top view of typical installation...
Page 21
Table 2 Product dimensions Product 5975 Series MSD • G3170A Diffusion EI MSD • G3171A Standard turbo EI MSD • G3172A Performance turbo EI MSD • G3174A EI/PCI/NCI MSD 7890 Series GC 6890 Series GC 6850 Series GC Data System Be sure to allow space for your carrier and reagent gases.
Agilent Technologies is not responsible for connecting to or establishing communication NO TE with your site LAN network. The representative will test the 5975 Series MSD’s ability to communicate on a mini-hub or -switch only. The IP addresses assigned to the instrument(s) must be fixed (permanently assigned) NO TE addresses.
Electrical Requirements You are responsible for providing appropriate electrical power and power outlets for all of the components in your 5975 Series MSD system. Power considerations include: • Voltage ranges of major components • Power configurations • Power requirements • Power plugs and cords...
Site Preparation Power configurations Electrical power for the MSD must be single-phase. The neutral wire cannot be used for safety grounding. The ground wire should carry zero current except for ground-fault current or static electric discharge. The entire system should share an isolated, noise-free electrical ground.
Power requirements Table 4 lists the power requirements for the 5975 Series MSD and related equipment. Extra power capacity for future additions is recommended. Each product listed requires a dedicated circuit. The gauge controller is powered by the MSD. The MSD, GC, and data system must each have a separate circuit breaker.
The MSD is supplied with a power cord and plug appropriate for the country from which the order originates. For example, if an order originates from a Agilent Technologies sales office in Germany, the power cord and plug supplied will be compatible with the standard voltage and outlet configuration in Germany.
Do not use extension cords with the MSD or GC. Extension cords normally are not WA RN ING rated to carry enough power and can be a safety hazard. Maintain easy access to the power cords so they can be disconnected for maintenance. NO TE Other electrical considerations Additional electrical considerations include:...
Each of these is considered in more detail in the following material. Temperature, humidity, and altitude The 5975 Series MSD is specified for operation under the following conditions: • Operation requires constant temperature (variations <2 °C/h) • Operation and storage require a noncondensing, noncorrosive atmosphere •...
Airborne dust Agilent Technologies recommends a maximum airborne particle density of 55 μg/m Technologies Customer Service Organization. Customer Engineers with special training and equipment can test for airborne particle density. They can offer suggestions for reducing airborne dust. Exhaust venting There are two sources of exhaust on the GC/MSD system: the foreline pump and the GC split vent.
Site Preparation See your GC manuals for information on venting the split vent. Fume (exhaust) hood An auxiliary work space and fume hood are needed for some maintenance procedures. Site Preparation Guide...
Carrier and Reagent Gas Requirements Table 6 lists the specifications for the carrier and reagent gases. Pressures and flows are as measured at the bulkhead fittings on the back of the instrument, not at the gas supply. We recommend that additional shut-off valves be installed near the instruments.
• They must supply gas at the pressure specified. • They must have one outlet with 1/4-inch Swagelok fittings. See the Agilent Technologies Chemical Analysis Columns and Supplies Catalog or visit the Agilent Technologies Chemical Analysis website at http://www.agilent.com/chem for dual-stage regulators available from Agilent Technologies.
Laboratory Supply Requirements Laboratory supply requirements vary with applications. Cleaning solvents Cleaning tasks for the MSD require the following HPLC-grade (or better) solvents: • Methylene chloride • Isopropyl alcohol • Methanol • Acetone • Water Proper storage, handling, and disposal of these chemicals is required for personal and environmental safety.
Spare parts and consumables The supplies and parts listed in Table 7 are used in the operation and maintenance of a 5975 Series MSD system. Keeping these parts on hand can reduce system downtime related to instrument maintenance and repair.
Page 35
Table 7 Chemical Consumables PFTBA (EI calibration fluid) PFDTD (CI calibration fluid) EVAL A (evaluation sample) OFN (Octafluoronapthalene) OFN 100 fg/µL Benzophenone PFHT Foreline pump oil, P3, 0.5 liter Gas Filters He/H gas purifier for GC Methane/Isobutane gas purifier for CI MSD Commonly used tools Safety glasses Wrenches...
Page 36
Site Preparation Table 7 0.5-mm id, 85% Vespel 15% graphite, for 0.32-mm id columns 0.8-mm id, 85% Vespel 15% graphite, for 0.53-mm id columns GC inlet 0.27-mm id, 90% Vespel 10% graphite, for 0.10-mm id columns 0.37-mm id, 90% Vespel 10% graphite, for 0.20-mm id columns 0.40-mm id, 90% Vespel 10% graphite, for 0.25-mm id columns...
When your MSD system is delivered, it is your responsibility to provide for removal of the shipping containers from the truck and their storage until installation. Contact your Agilent Technologies service representative as soon as your shipment arrives to arrange an installation date.
Do not open any shipping containers until an Agilent Technologies representative is present. Warranty claims for missing items will not be honored unless an Agilent Technologies representative is present to verify the contents of each container as it is unpacked.
Further, the Agilent Technologies service representative will not set up your laboratory procedures. Assistance with laboratory procedures can be obtained from your local Agilent Technologies Applications Engineer (AE) on a consulting basis at additional cost. Site Preparation Guide...
Site Preparation Sensitivity specifications See the appropriate sensitivity specifications for your instrument type. See the Agilent Technologies Web site at http://www.agilent.com/chem. SIM performance verification is not performed at installation and must be purchased if NO TE desired. Site Preparation Guide...
Page 41
5975 Series Mass Selective Detectors Site Preparation Guide Power Cords This appendix shows the power cords available for the 5975 Series MSD. See page 26 for information about ensuring that the correct power cord is supplied with your system. US and Canada, NEMA 6-15P (Part number 8120-6825)
Page 42
Power Cords Denmark/Greenland, IEC 309 (Part number 8120-3997) Australia/New Zealand, AS 3112-1981 (Part number 8120-1369) India/South Africa, BS 546 (Part number 8120-4211) Israel, SI 32 (Part number 8120-5182) Japan, NEMA L6-20P (Part number G2025-60189) Site Preparation Guide...