2 - 8
Fig. 5
Check the turn plate
?
•
Push one of the turn plate buttons. The buttons must be lit when the turn plate is
activated.
•
Check that the unit can rotate around the turn plate, see fig. 5.
•
Push the turn plate button again and check that the turn plate goes back in parking
position.
•
Repeat the check with the other turn plate button. If any problem occurs, see "Service
Instructions, section Adjustment and replacement of the turn plate".
Check the brake
?
•
Lock the brake. It should not be possible to move the MOBILETT with normal hand effort.
A slight skip is acceptable when using a force, greater than 350N.
Check the light field to x-ray field
?
•
Load a 35x35 cm (14"x14") cassette with film.
•
Put the cassette on a table or a similar repository.
•
Set a SID of 100 cm (39"). Use the measuring tape.
•
Switch on the light localizer and align the cassette. Now, collimate the light field to
25 x 25 cm (10"x10").
•
Mark the four corners with coins and mark one of the corners with two coins, see fig. 6.
•
Set 52 kV, 2.0 mAs on the control panel.
•
Trigger an exposure.
MOBILETT Plus HP
Installation and setting
Fig. 6
SPR8-220.033.01
Rev. 05
05.01
Page 8 of 10
Siemens-Elema AB
SPS-UD
2
2
2