Test of kV-accuracy

?
Requirement:
- The measured kV values, shall be within the limits.
Required measurement devices:
- kV-meter, based on the filter-comparison method, for example PTW-Nomex kV meter.
When measuring the kV non invasive a correction factor from the user manual of the
kV-meter has to be used. If a kV meter is not available and if it is within local laws and
regulations, the measurement can be made with an oscilloscope on D1. KVPEAK - D1
GND with a result of 40 kV/V.
NOTICE
For the actual filtration of this MOBILETT, see the test protocol.
If the actual inherent filtration is unknown, the correction shall be made with a reference
value of 5 mm Al.
Put the kV meter into operation, see operating instructions for the kV-meter.
Place the measurement detector on a table or other. Using the light field, set the detector
in the beam projection, see fig. 4.
Set the SID to approx. 50 cm (20"). The SID may vary between different kV-meters.
Fig. 4
MOBILETT Plus HP
Higher filtration results in higher kV values.