User-Entered Calibration,
Note: For meters with factory-calibrated ranges or methods, Standard Calibration Adjust
(SCA) will be disabled when a user-entered method is programmed into the meter. To
turn SCA back on, restore the meter to factory default calibration.

Maximum/Minimum Displayed Value

In meters with absorbance (Abs) ranges, the maximum displayed value and
minimum displayed value is related to the value of the standards entered in a user
calibration.
Measurements that exceed the minimum or maximum standards entered in the
user calibration will return a flashing number indicating "underrange" or
"overrange". See Error Codes (page 2—12) for more information.
Example 1
For a calibration with the following standards:
S0=0.000
S1=1.000
Maximum Displayed Value
Minimum Displayed Value
网址:www.mamots.com
continued
1.000
0.000
2—26
客服:400 688 9809
电话:0431-85178766