Calibration Procedure - Hach PHC745 User Manual
Hide thumbs
1
2
Table Of Contents
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
page
of
26
Go
/
26
Contents
Table of Contents
Troubleshooting
Bookmarks
Table of Contents
Advertisement
5.2 Calibration procedure
1. Go to the
calibrate menu.
Select the probe, if
applicable. The
display shows the
pH buffer solutions
to use for calibration.
5. Loosen the
sleeve junction. Put
the probe in the first
pH buffer solution.
Make sure that the
sensor and
reference junction
are fully in the
solution. Do not put
the probe on the
bottom or sides of
the beaker.
9. Save the
calibration.
2. Prepare or pour
the pH buffer
solutions in different
beakers.
6. Shake the probe
from side to side to
refresh the reference
junction and remove
air bubbles.
Note: Be careful to not
hit the beaker and
damage the probe.
3. Open the filling
hole.
7. Stir slowly, then
read the pH value of
the buffer solution.
The display shows
the temperature-
corrected pH value
when the reading is
stable.
4. Rinse the probe
with deionized water.
Dry the probe with a
lint-free cloth.
8. Continue with
steps
4
through
7
for
the remaining
buffers or select
Done.
English 9
Table of
Contents
Previous
Page
Next
Page
1
...
8
9
10
11
12
13
14
15
Advertisement
Table of Contents
Related Manuals for Hach PHC745
Laboratory Equipment Hach sensION+ pH3 User Manual
(15 pages)
Laboratory Equipment Hach sension ph31 User Manual
(19 pages)
Laboratory Equipment Hach PHOSPHAX sc LR User Manual
(64 pages)
Laboratory Equipment Hach PHOSPHAX sigma High Range User Manual
(36 pages)
Laboratory Equipment Hach PHC301 User Manual
(20 pages)
Laboratory Equipment Hach PHC281 User Manual
(20 pages)
Laboratory Equipment Hach PHC108 User Manual
(18 pages)
Laboratory Equipment Hach PHC729 User Manual
(22 pages)
Laboratory Equipment Hach PHC735 User Manual
(22 pages)
Laboratory Equipment Hach PHC201 User Manual
(18 pages)
Laboratory Equipment Hach PHC705 User Manual
(22 pages)
Laboratory Equipment Hach POCKET TURBIDIMETER User Manual
Analysis system (56 pages)
Laboratory Equipment Hach POCKET COLORIMETER II Instruction Manual
Analysis systems (94 pages)
Laboratory Equipment Hach PTM450 User Manual
(164 pages)
Laboratory Equipment Hach SIGMA 900 Instrument Manual
Composite sampler (52 pages)
Laboratory Equipment Hach 900 MAX Instrument Manual
Refrigerated sampler (160 pages)
Related Content for Hach PHC745
2100AN Stablcal Calibration Procedure
Hach 2100AN
PHC705 Calibration Procedure
Hach PHC705
PHC201 Calibration Procedure
Hach PHC201
POCKET COLORIMETER II Calibration Procedure Using Prepared Standards
Hach POCKET COLORIMETER II
PHC735 Calibration Procedure
Hach PHC735
PHC729 Calibration Procedure
Hach PHC729
ISENH4181 Calibration Procedure
Hach ISENH4181
ISEF121 Series Calibration Procedure
Hach ISEF121 Series
PHC108 Calibration Procedure
Hach PHC108
PHC281 Calibration Procedure
Hach PHC281
PHC301 Calibration Procedure
Hach PHC301
ISECL181 Calibration Procedure
Hach ISECL181
2100AN Using Color Units (Platinum Cobalt Color Units Calibration Procedure)
Hach 2100AN
sension ph31 Calibration Procedure
Hach sension ph31
POCKET TURBIDIMETER Calibration Procedure
Hach POCKET TURBIDIMETER
2100AN IS Stablcal Calibration Procedure
Hach 2100AN IS
Table of Contents
Print
Rename the bookmark
Delete bookmark?
Delete from my manuals?
Login
Sign In
OR
Sign in with Facebook
Sign in with Google
Upload manual
Upload from disk
Upload from URL