Omega CL-355A User Manual

Omega CL-355A User Manual

Dry-block calibrator

Advertisement

Quick Links

User' s Guide
®
Shop online at
omega.com
CL-355A
e-mail: info@omega.com
For latest product manuals:
Dry-Block Calibrator
omegamanual.info
MADE IN UK

Advertisement

Table of Contents
loading

Summary of Contents for Omega CL-355A

  • Page 1 User’ s Guide ® Shop online at omega.com CL-355A e-mail: info@omega.com For latest product manuals: Dry-Block Calibrator omegamanual.info MADE IN UK...
  • Page 3: Specification

    Specification Figures quoted are at the base of the well at the time of calibration. Temperature range: 5°C/9°F above ambient to 400°C/752°F Over-temperature limit: 450°C/842°F (approximately) Display resolution: 0.1° Accuracy: +/-0.6°C (1.1°F) Stability: +/-0.15°C (0.27°F) Heat up time 20° C to 400°C: 12 minutes Cool down 400°C to 100°C: 21 minutes...
  • Page 4: Working Environment

    If the equipment is used in a manner not specified by Omega, the protection provided by the equipment to the operator may be impaired. All Omega Engineering units have been designed to conform to international safety requirements and are fitted with a preset over-temperature cutout.
  • Page 5: Installation

    Installation 1. All Omega units are supplied with a power cable. 2. Before connecting the mains supply, check the voltage against the rating plate. Connect the mains cable to a suitable plug according to the table below. Note that the unit must be earth grounded to ensure proper electrical safety.
  • Page 6: Operation

    OPERATION Preparation The heater design, temperature sensor and control circuit give good temperature control and uniformity, but make sure that there is a close fit of the probes in the block to allow efficient heat transfer. Plug the power cable into the socket in the back of the unit. Connect the mains cable to the electricity supply.
  • Page 8: Setting The Operating Temperature

    Setting the operating temperature 1. To set the temperature required, press and hold either the up or down button depending on the direction you need to take the set point. 2. When you have the correct set temperature displayed, release the up or down button and the unit will start to heat or cool to the set point.
  • Page 9: Operator Maintenance

    Please include any details of the fault observed and remember to return the unit in its original packing. Omega will accept no responsibility for any damage to units that are improperly packed for shipment. If in doubt, contact your supplier. See the Decontamination Certificate supplied with your unit.
  • Page 10: Replacement Parts

    Replacement Parts The following parts may be obtained from Omega Engineering if replacements or alternatives are required: Part Number Description FCABLEUK UK 240 volt Mains cable with 13amp UK plug (5 amp fuse) FCABLEEU Euro style 240 volt Mains cable with R/A Schuko plug...
  • Page 11: Warranty Disclaimer

    FAX: +31 20 643 46 43 e-mail: sales@omegaeng.nl CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic Component” under 10 Czech Republic: Frystatska 184 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any 733 01 Karviná, Czech Republic...

Table of Contents