Using The Appliance; Temperature Regulation - Dometic RA 122 H Operating Instructions Manual

Health care refrigerator
Table of Contents

Advertisement

3. Ventilation must be provided as shown in
alternatives A, B, C or D (fig 2-3)
4. The ventilation duct must at least measure
105mm x the width of the refrigerator
5. Only the entire cooling unit must project into
the duct as shown
6. Air passing through the duct must not be
preheated by any source of heat
7. Ventilation grills, if used, must have openings
of at least 200cm 2 each
During installation, please ensure the mains plug
is accessible at all times.

5. Using the appliance

To start the refrigerator connect the plug to the socket
outlet and switch on the socket. The cooling fins
should show signs of cooling after about 3 hours.
The refrigerator needs about 6 hours at normal
ambient temperature to reach 6 °C in the storage
compartment.

5.1. Temperature regulation

The usual setting required is in the region of 4 to 5 on
the thermostat knob (Fig. 4), but this can vary with
conditions of load and installation. If it is found
necessary to alter the setting,the knob can be turned
to a higher number to produce a colder temperature,
or a lower number to produce a warmer temperature.
The effect of any adjustment can be checked by
means of the
thermometer after leaving the
refrigerator operating for a further period to settle
down again.