8 Illustrated parts

8.1 Special instructions

8.2 Service tools

8.2.1 Calibration
accessories kit, ST 2
8.2.2 Test lung kit ST 3
8.2.3 "Service" flow
screen ST 5
8-2
Apply a thin coat of oxygen-use-approved lubricant to o-rings prior to installation
(unless otherwise noted).
When replacing fittings, position the barb end in the same direction as the original
fitting to make the routing and connection of tubing easier.
The calibration accessories kit (P/N MA-640008) consists of
2 x 22 mm corrugated tube, 15"
22 mm sealer.
The test lung kit (P/N AZ-381501) consist of
Test lung with built in resistor, Vt range = 50 – 500 ml
R approx. 15 – 25 mbar/l/s, , C approx. 10 - 30 ml/mbar
Filter
Note
This test lung kit is used for servicing the Centiva/5 Plus ventilator. Test lungs with
different compliance and resistance characteristics may not match test lung
related data in this service manual.
Note
A minimum resistance of 5 mbar/l/s is required.
The "Service" flow screen (P/N CX-526165) is a flow screen with tolerance "0%",
used for calibration of the expiratory flow sensor according to chapter 4.7.5.The
flow screen comes in a plastic protection case.
TRM, Centiva/5 Plus, M1047448/ CM-922026, rev. 3.50.a