Safety precautions
1�3�6
Precautions regarding
maintenance
WARNING
• Perform the maintenance and inspection periodically. In
addition to the user periodical maintenance that needs
to be performed, periodical maintenance by a service
engineer is also required.
• If there are stains such as body fluids, clean and disin-
fect
• Based on the warranty, parts that are no longer under
warranty (1 year) can be replaced for a fee.
• Turn off the power and disconnect the power plug from
the wall outlet before cleaning or maintaining this de-
vice.
• Securely connect the power cable, wired cable and
Ethernet cable after cleaning and maintenance.
• Clean the dirt from the gaps between the exterior of DR
Detector, and the terminal using a commercial plastic
brush. Do not clean with sharp or hard objects.
• Take care regarding the following when disinfecting the
DR Detector.
— Use ethanol for disinfection, isopropanol for disinfec-
tion, or commercial chlorine bleach, or 0.5 % hypo-
chlorite (10-fold dilution of household bleach) when
disinfecting. However, bleach and hypochlorite are
corrosive, so wash the bleach off well to avoid corro-
sion
— Please note that if you use chemical for disinfection
other than the above, it may affect the quality, perfor-
mance, and safety of the DR Detector.
— If you directly apply or spray disinfecting solution on
the DR Detector, the solution will enter the instrument
through exterior gaps, causing a failure. Dampen a
lint-free, soft cloth with disinfecting solution, and use
after wringing it thoroughly. Do not apply disinfecting
solution onto the wired connection connector and
LED when cleaning.
— Disinfecting solution is a chemical agent, so follow
the precautions of the manufacturer.
• Fully charge battery once a month even if the DR De-
tector has not been used for a long time.
CAUTION
1�3�7

Precautions on service life

Service life
Name
DR Detector
AeroDR Interface Unit
AeroDR Interface Unit2
Detector Interface Unit
Detector Interface Unit 2
Power Supply Unit
AeroDR Generator Interface Unit
AeroDR Generator Interface Unit2
Generator Interface Unit 3
AeroDR Battery Charger
AeroDR Battery Charger2
AeroDR Access Point
I/F Cable
AeroDR UF Cable
• The above service life is valid only if the product has
been properly operated while following the precautions
for use and performing the specified maintenance. (By
self certification <our data>)
• The service life may differ depending on usage condi-
tions and environment.
• Some component parts of this device are commercially
available parts that have a short cycle of model chang-
es, therefore, it might not be possible to supply service
parts even within the service life. In addition, related
component parts may need to be replaced to maintain
compatibility at the time of model change.
28
CAUTION
Service life
6 years
6 years
6 years
6 years
6 years
6 years
6 years
6 years
6 years
6 years
6 years
6 years
6 years
6 years