Agilent Technologies Cary 600 Series User Manual

Agilent Technologies Cary 600 Series User Manual

Ftir spectrometers and microscopes

Advertisement

Agilent Cary
600 Series FTIR
Spectrometers and
Microscopes
User's Guide

Advertisement

Table of Contents
loading

Summary of Contents for Agilent Technologies Cary 600 Series

  • Page 1 Agilent Cary 600 Series FTIR Spectrometers and Microscopes User’s Guide...
  • Page 2 DFAR 252.227-7014 (June 1995), or as a “commercial item” as defined in FAR 2.101(a) or as “Restricted computer software” as defined in FAR 52.227-19 (June 1987) or any equivalent agency regulation or Agilent Cary 600 Series FTIR Spectrometers and Microscopes User’s Guide...
  • Page 3: Table Of Contents

    Storage Electrical hazards Laser safety Labeling Laser aperture locations Other precautions Warning and caution messages Information symbols Color coding CE compliance Electromagnetic compatibility EN55011/CISPR11 ICES/NMB-001 2. Introduction Specifications Environmental Fuses Agilent Cary 600 Series FTIR Spectrometers and Microscopes User’s Guide...
  • Page 4 Programming accessories for hardware recognition 4. Operation Lights Turning on the system and filling the Dewar Spectrometer Microscope Performing a collect Spectrometer Microscope Accessing the Help Turning off the system Agilent Cary 600 Series FTIR Spectrometers and Microscopes User’s Guide...
  • Page 5 Contents 5. Maintenance Routine Cleaning Spare parts 6. Troubleshooting Agilent Cary 600 Series FTIR Spectrometers and Microscopes User’s Guide...
  • Page 6 Contents This page is intentionally left blank. Agilent Cary 600 Series FTIR Spectrometers and Microscopes User’s Guide...
  • Page 7: Safety Practices And Hazards

    (both hard copy and online) provided with your instrument and accessories. Before using the instrument or accessories, you must thoroughly read these safety practices. Observe all relevant safety practices at all times. Agilent Cary 600 Series FTIR Spectrometers and Microscopes User’s Guide...
  • Page 8: General

    Safety Practices and Hazards General Operation of an Agilent Cary 600 Series FTIR spectrometer and microscope involves the use of compressed gases, high voltage energy, and hazardous materials including liquid nitrogen, tungsten halogen source lamps that emit ultraviolet radiation and a helium- neon laser operating in the visible region at 632.8 nanometers.
  • Page 9: Compressed Gas Cylinders

    Safety Practices and Hazards Compressed gas cylinders Consult the Agilent Cary 600 Series FTIR Site Preparation Guide for gas pressure requirements. Compressed gas cylinders contain highly pressurized gas. If storage conditions are outside of the recommended supplier’s safety codes, the cylinders can explode or rapidly release gas into the environment.
  • Page 10: Storage

    Application of the wrong supply voltage can create a fire hazard and a potentially serious shock hazard, and could seriously damage the Cary 600 Series FTIR system, accessories and any attached ancillary equipment. The Cary 600 Series FTIR instruments have a universal power supply that adapts to the supply voltage.
  • Page 11: Laser Safety

    632.8 nanometers. The spectrometer is a Class 2 laser product, powerful enough to warrant caution in its use. Agilent Cary 600 Series FTIR spectrometers and microscopes comply with FDA and CE standards for light emitting products.
  • Page 12: Labeling

    Laser safety labels are located on the rear of the Agilent Cary 600 Series FTIR spectrometer and underneath the top cover. Figure 1. Laser safety label locations on the rear of the Agilent Cary 600 Series FTIR spectrometer The laser safety warning just above the Purge and Air Bearing connections states ‘Laser radiation.
  • Page 13 The following image shows the laser safety warning located near all apertures on both the Agilent Cary 600 FTIR Series instruments and microscopes. Figure 3. Laser safety warning located near apertures Agilent Cary 600 Series FTIR Spectrometers and Microscopes User’s Guide...
  • Page 14: Laser Aperture Locations

    Infrared sources operate at high temperatures, which may burn you. Before replacing a source element that has been lit, switch off the instrument and ensure that the source has adequately cooled. Agilent Cary 600 Series FTIR Spectrometers and Microscopes User’s Guide...
  • Page 15: Warning And Caution Messages

    Consult the manuals supplied with your computer, monitor and printer for their specific ventilation requirements. Use of the Agilent Cary 600 Series FTIR system and accessories may involve materials, solvents and solutions which are flammable, corrosive, toxic or otherwise hazardous.
  • Page 16: Information Symbols

    The following symbol may be used on warning labels attached to the instrument. When you see this symbol, refer to the relevant operation or service manual for the correct procedure referred to by that warning label. Agilent Cary 600 Series FTIR Spectrometers and Microscopes User’s Guide...
  • Page 17: Color Coding

    A green light indicates the instrument is in normal or standby  mode. A yellow light indicates that the instrument needs attention.  A blue light indicates that operator intervention is required.  Agilent Cary 600 Series FTIR Spectrometers and Microscopes User’s Guide...
  • Page 18: Ce Compliance

    This device complies with the requirements of CISPR11, Group 1, Class A as radiation professional equipment. Therefore, there may be potential difficulties in ensuring electromagnetic compatibility in other environments, due to conducted as well as radiated disturbances. Agilent Cary 600 Series FTIR Spectrometers and Microscopes User’s Guide...
  • Page 19: Ices/Nmb-001

    Technologies could void the user’s authority to operate the equipment. ICES/NMB-001 This ISM device complies with Canadian ICES- 001. Cet appareil ISM est conforme à la norme NMB-001 du Canada. Agilent Cary 600 Series FTIR Spectrometers and Microscopes User’s Guide...
  • Page 20 Safety Practices and Hazards This page is intentionally left blank. Agilent Cary 600 Series FTIR Spectrometers and Microscopes User’s Guide...
  • Page 21: Introduction

    Pollution degree 2 and Installation Category II (EN 61010-1). All specifications are included in the Agilent Cary 600 Series FTIR Site Preparation Guide. Agilent Cary 600 Series FTIR Spectrometers and Microscopes User’s Guide...
  • Page 22: Environmental

    Introduction Environmental For proper operation of your Agilent Cary 600 Series FTIR spectrometers and microscopes, these conditions are required: Temperature: 20 °C to 26 °C (68 °F to 80 °F)  Temperature gradient: 1 °C/hr (1.8 °F/hr) maximum  Relative humidity: 20% to 50% non-condensing ...
  • Page 23: Purge Gas

    Specifications for the embedded laser in the spectrometer are: Maximum accessible power: < 1 mW  Pulse duration: Continuous  Wavelength: 632.8 nm  Maximum beam divergence: 1.3 milliradians  Agilent Cary 600 Series FTIR Spectrometers and Microscopes User’s Guide...
  • Page 24: Liquid Nitrogen

    200 milliliters (7 ounces) of liquid nitrogen to top off the Dewar. See Page 36, for more information on filling the spectrometer Dewar and Page 38 for more information on filling the microscope Dewar. Agilent Cary 600 Series FTIR Spectrometers and Microscopes User’s Guide...
  • Page 25: Installation Requirements

    Cary 600 Series FTIR system. You must prepare your laboratory according to these instructions before the Cary 600 Series FTIR can be installed. You should keep the Site Preparation Guide for future reference. If you have misplaced your copy, you can obtain a replacement from your local Agilent office.
  • Page 26: Documentation

    Bolded items indicate an action. For example, ‘click OK’ and  ‘From the Edit menu, choose Copy’. ALL CAPITALS indicate keyboard commands. For example,  ‘press ENTER’ and ‘press SHIFT+F8’. Agilent Cary 600 Series FTIR Spectrometers and Microscopes User’s Guide...
  • Page 27: Unpacking And Installation

    Agilent and the carrier within five days. Do not accept packages with obvious puncture damage. All other damage should be noted on the shipping document and signed by the delivery agent. Agilent Cary 600 Series FTIR Spectrometers and Microscopes User’s Guide...
  • Page 28: Installing The Software

    CAUTION microscope unless otherwise instructed by your Agilent representative. Your Agilent representative will complete the unpacking and installation of your Agilent Cary 600 Series FTIR spectrometers and microscopes. If incorrectly unpacked, equipment may become damaged. Installing the software While your Agilent representative will install the Resolutions Pro software for you during the installation process, it may become necessary to re-install it if the computer is replaced.
  • Page 29 Figure 6. Purge fittings on the rear of the spectrometer CAUTION Use only dry air or dry nitrogen to purge the spectrometer and/or microscope Agilent Cary 600 Series FTIR Spectrometers and Microscopes User’s Guide...
  • Page 30 10 USER 5 IN - Used to receive a 11 Beam splitter access cover 12 Metal cover for cable routing signal from an external device for second ADC (if fitted) 13 Sample compartment connectors (not shown) Agilent Cary 600 Series FTIR Spectrometers and Microscopes User’s Guide...
  • Page 31: Microscope

    For purging of the microscope, connect the gas purge line to the fitting on the rear of the microscope upper cover. CAUTION Use only dry air or dry nitrogen to purge the spectrometer and/or microscope Agilent Cary 600 Series FTIR Spectrometers and Microscopes User’s Guide...
  • Page 32: Connecting The Motorized Stage Cables

    EXT DET 1 or EXT DET 2 spectrometer penthouse connector. Connecting the motorized stage cables There are six connectors on the rear of the stage controller. Figure 11. Motorized stage controller connections Agilent Cary 600 Series FTIR Spectrometers and Microscopes User’s Guide...
  • Page 33 Connects to the X/Y stepping motors on the right side of the stage (see 1 on Figure 12) Figure 12. Motorized stage connections Where: X/Y stage control Z stage control Agilent Cary 600 Series FTIR Spectrometers and Microscopes User’s Guide...
  • Page 34: Programming Accessories For Hardware Recognition

    See the Resolutions Pro Help for instructions on how to program accessories. Specifically, click ‘Start’ > ‘All Programs’ > ‘Agilent Resolutions’ > ‘Resolutions Pro Help’ > ‘Accessories’ > ‘Programming Accessories for Hardware Recognition’. Agilent Cary 600 Series FTIR Spectrometers and Microscopes User’s Guide...
  • Page 35: Operation

    Click the plus (+) icon next to ‘Agilent 600 Series FTIR’ to expand it for hardware information on your spectrometer or click the plus (+) icon next to ‘610-IR’ for hardware information on your Agilent Cary 610-IR microscope. Agilent Cary 600 Series FTIR Spectrometers and Microscopes User’s Guide...
  • Page 36: Lights

    This procedure applies to detectors requiring liquid nitrogen. Most instruments CAUTION are supplied with DLaTGS detectors, which do not require nitrogen cooling. Always check the detector type and cooling requirements before performing this procedure. Agilent Cary 600 Series FTIR Spectrometers and Microscopes User’s Guide...
  • Page 37 Raise the spectrometer cover. The Dewar access hole is on the right rear of the detector compartment cover. Dewar fill hole Detector compartment cover – raise spectrometer cover to access. Figure 13. Dewar hole on spectrometer Agilent Cary 600 Series FTIR Spectrometers and Microscopes User’s Guide...
  • Page 38: Microscope

    To fill the microscope Dewar with liquid nitrogen: Turn on the microscope power switch located on the rear of the microscope. Remove the appropriate liquid nitrogen Dewar plug. This plug is easily removed by hand. Agilent Cary 600 Series FTIR Spectrometers and Microscopes User’s Guide...
  • Page 39 Allow about 20 minutes for the Dewar to come to thermal equilibrium. Top up the Dewar with about 200 milliliters (7 ounces) of liquid nitrogen. Replace the plug. Agilent Cary 600 Series FTIR Spectrometers and Microscopes User’s Guide...
  • Page 40: Performing A Collect

    For instructions on how to access the Help, which contains information for each of these steps, see Page 41. The shutter should always be in the open position when using the microscope. NOTE Agilent Cary 600 Series FTIR Spectrometers and Microscopes User’s Guide...
  • Page 41: Accessing The Help

    For collect information, highlight ‘Perform a Collect’ and click the links for the required information or highlight ‘Collecting Microscope Data’ or ‘Microscope Mapping Data’ for information on using the microscope to collect data. Agilent Cary 600 Series FTIR Spectrometers and Microscopes User’s Guide...
  • Page 42: Turning Off The System

    Using the computer’s power switch to quit Windows may cause data to be lost. CAUTION Always use the procedure advised by Microsoft. If necessary, turn off the spectrometer or microscope power switch. Agilent Cary 600 Series FTIR Spectrometers and Microscopes User’s Guide...
  • Page 43: Maintenance

    600 FTIR spectrometers and microscopes maintenance procedures that may be carried out by an operator. Any maintenance procedures not specifically mentioned should be carried out only by Agilent- trained, Agilent-qualified or Agilent-authorized representatives. Agilent Cary 600 Series FTIR Spectrometers and Microscopes User’s Guide...
  • Page 44 WARNING Contact with these circuits, devices and components can result in death, serious injury, or painful electrical shock. This instrument contains electrical circuits, devices and components operating at dangerous voltages. Agilent Cary 600 Series FTIR Spectrometers and Microscopes User’s Guide...
  • Page 45 This section refers only to maintenance procedures for the instrument. You NOTE should refer to your computer and printer manuals for their maintenance procedures and to Resolutions Pro Help for the maintenance procedures for any accessories you ordered. Agilent Cary 600 Series FTIR Spectrometers and Microscopes User’s Guide...
  • Page 46: Routine

    Do not use organic solvents or abrasive cleaning agents. Spare parts For spare parts and consumables ordering information, refer to the Agilent website, www.agilent.com Agilent Cary 600 Series FTIR Spectrometers and Microscopes User’s Guide...
  • Page 47: Troubleshooting

    Start the Resolutions Pro Software. Press F1 on the keyboard to display the Help. Click the Contents tab if necessary. Click the plus (+) icon next to ‘Troubleshooting’ to expand it. Agilent Cary 600 Series FTIR Spectrometers and Microscopes User’s Guide...
  • Page 48 Troubleshooting This page is intentionally left blank. Agilent Cary 600 Series FTIR Spectrometers and Microscopes User’s Guide...
  • Page 50 In This Guide The guide describes the following: Safety Practices and Hazards  Introduction  Unpacking and Installation  Operation  Maintenance  Troubleshooting  © Agilent Technologies 1997, 2000- 2004, 2010-2013 Printed in Malaysia 05/13 *8510246800* *8510246800* 8510246800 Issue 10...

This manual is also suitable for:

Cary 610

Table of Contents