Zeiss S8 Instructions For Use Manual

Floor stand
Hide thumbs Also See for S8:

Advertisement

S8 Floor Stand
Instructions for use
G-30-1346-en
Issue 4.0
Printed on 05. 06. 2001

Advertisement

Table of Contents
loading

Summary of Contents for Zeiss S8

  • Page 1 S8 Floor Stand Instructions for use G-30-1346-en Issue 4.0 Printed on 05. 06. 2001...
  • Page 3: Table Of Contents

    Contents Functions at a glance – S8 Floor Stand – Key to symbols Safety – Directives and standards – Notes on installation and use – Safety devices – Warning labels and notes Description S8 floor stand – Intended use – Description of the modules –...
  • Page 4 – Changing the xenon lamp module – Care of the unit – Sterilization – Disinfecting the control keys – Ordering data – Accessories – Spare parts – Technical data – Ambient requirements Index G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 5: Functions At A Glance

    Functions at a glance Functions at a glance S8 Floor Stand Key to symbols G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 6: S8 Floor Stand

    Functions at a glance S8 Floor Stand Halogen illumination system 1 Closed flap: main lamp is on - Open flap: backup lamp is on 2 Filter knob 3 Opening the lamp module 4 Manual selection of backup lamp Xenon illumination system...
  • Page 7 Functions at a glance G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 8: Key To Symbols

    Note: The hand indicates hints on the use of the instrument or other tips for the user. ® OPMI ® OPMI is a registered trademark of Carl Zeiss. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 9: Safety

    Safety Safety Directives and standards Notes on installation and use Safety devices Warning labels and notes G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 10: Directives And Standards

    Safety The instrument described in this manual has been developed and tested in accordance with Carl Zeiss safety standards and with national and in- ternational regulations. A high degree of instrument safety is thus en- sured. We would like to inform you on the safety aspects involved in operating the instrument.
  • Page 11: Notes On Installation And Use

    Use this instrument only for the applications described. • Only use the instrument with the accessories supplied. Should you wish to use other accessory equipment, make sure that Carl Zeiss or the equipment manufacturer has certified that its use will not impair the safety of instrument.
  • Page 12 Make sure that all ”Requirements for operation” are fulfilled. • Go through the checklist. • Re-attach or close any covers, panels or caps which have been re- moved or opened. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 13 After every use of the instrument • Always use the main power switch of the instrument to turn it off. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 14 Safety • The main power switch must always be turned off when the instrument is not in use. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 15 Safety G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 16: Safety Devices

    When locked, the suspension arm can no longer suddenly spring up- ward when insufficient weight is attached. After attaching a unit, per- form the balancing procedure. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 17 Safety G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 18 – Blinks when the backup lamp has failed. 4 Manual function When the manual function has been activated, all electrical control systems are inoperative. The lamp brightness is automatically ad- justed to a fixed setting. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 19 Safety G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 20 3 Manual function When the manual function has been activated, all electrical control systems are inoperative. The lamp brightness is automatically ad- justed to a fixed setting. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 21 Safety G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 22 Press the Manual key once again to reactivate electronic control; the display in the display and key field then returns to the basic mode. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 23 Safety G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 24: Warning Labels And Notes

    Observe all warning labels and notes! If any label is missing on your instrument or has become illegible, please contact us or one of our authorized representatives. We will supply the missing labels. OPHTHALMOLOGY G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 25 Safety G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 26 Safety G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 27: Description

    Description Description S8 floor stand Intended use Description of the modules Design Stand base with stand column Connection panel Suspension arm Display field with control keys Halogen illumination system Xenon illumination system G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 28 Description G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 29: S8 Floor Stand

    The S8 floor stand is a suspension system for Zeiss surgical microscopes. It is used to power and control the motorized functions of the surgical mi- croscope. The hallmarks of the S8 floor stand are its superb mobility and easy operation. The motorized functions can be controlled using a foot control panel or hand control panel.
  • Page 30: Description Of The Modules

    The base is easy to move on its four casters. The base has been designed in such a way that high stability is guaranteed even with unfavorable loading of the stand. A brake allows the S8 floor stand to be quickly and reliably secured in position.
  • Page 31 Description G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 32 Description The S8 floor stand is optionally equipped with two different types of illumi- nation system. 1 Halogen illumination The halogen illumination comprises two separate systems for fiber il- lumination. Each of the two lamp housings contains a backup lamp which is automatically swung into the illumination beam path when the first lamp fails.
  • Page 33 Description G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 34: Stand Base With Stand Column

    3 Brake Press once to lock the stand in position. Press again to release the brake. 4 Steerable castors 5 Wheels G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 35 Description G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 36: Connection Panel

    The strain relief device prevents inadvertent unplugging of the fol- lowing electrical connections: – power cable – connecting cable for foot control panel, hand control panel or op- erating chair with an appropriate footswitch. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 37 Description G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 38: Suspension Arm

    After pressing any one of the keys, you can move the articu- lated arm as required. When you release the key, the magnetic clutches lock all axes simultaneously. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 39 Description G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 40: Display Field With Control Keys

    (modes). "MODE" key and "STORE" key The "MODE" and "STORE" keys permit you to access the different modes of the user interface. For details, please see the chapter "Operation". G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 41 The illustration shows the display and key field of the ceiling mount (floor stand) with two halogen illumination systems. This configuration is re- ® quired, for example, for OPMI VISU 200. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 42: Halogen Illumination System

    The filter knobs have four positions: no filter GG 475 filter: to protect the patient's eye during surgery against unnecessary (blue) radiation (retinal injury). KK 40 filter: to increase color temperature no filter G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 43 Description G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 44 – that one illumination system each can be switched on the left-hand and right-hand side of the foot control panel, – or that both illumination systems can be switched on the left-hand or right-hand side of the foot control panel. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 45 Description G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 46: Xenon Illumination System

    Turn knob (2) through 180° until it snaps in. This moves the backup lamp into the illumination beam path. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 47 Description G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 48 Note: You can adjust the selector in such a way that you can switch the illu- mination on/off on the right-hand and left-hand sides of the foot control panel. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 49 Description G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 50 Description G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 51: Preparations For Use

    Preparations for use Preparations for use Attaching the equipment Connections Connecting the surgical microscope Connecting the S light guide Mounting the strain relief device Connecting the stand Positioning the stand G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 52: Attaching The Equipment

    Slide the coupling (2) from above on the microscope shaft (8). Screw in the mounting screw (5) from above and tighten the securing screw (4) firmly using a 4 mm Allen key. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 53 Preparations for use G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 54 Before using and after re-equipping the unit, always make sure that se- curing screws (3), (4) and (7) have been tightened firmly. Note: When mounting surgical microscopes with an integrated coupling, steps 3 to 9 need not be performed. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 55 Preparations for use G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 56: Connections

    (7) of the microscope, and press the light guide into the cable clip (6). Note: Make sure that the light guide is not stretched or kinked when the micro- scope is turned or tilted. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 57 Preparations for use Surgical microscope shown rotated G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 58: Mounting The Strain Relief Device

    Open flap (4). • Feed the cable through opening (5). • Close flap (6). • Tighten the cable until it encloses flap (7). • Check the length of the cable. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 59 Preparations for use 320 mm G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 60: Connecting The Stand

    Connect the stand to line power using the power cord supplied for this purpose. Only use power outlets which are provided with a properly con- nected protective earth connector. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 61: Positioning The Stand

    Please also read the section " Relocating the stand ". Caution: Press the brake on the stand base and make sure that the stand is se- curely locked in position and cannot roll away. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 62 Preparations for use G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 63: Operation

    Operating the OPMI® PRO magis What to do in an emergency Failure of a halogen lamp Failure of the xenon lamp Failure of lamp control Failure of focusing system Failure of magnetic clutches G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 64: Preparations For Use

    20 cm. Using adjustment screw (2), adjust the spring force in such a way that the effort required to move the arm up or down is the same in both directions. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 65 Operation G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 66: Adjusting The Limit Of Downward Movement

    • Turn the adjustment screw (1) clockwise as far as it will go. • Move the surgical microscope again to the lower limit and check the safety distance. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 67 Operation G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 68: Relocating The Stand

    Be extremely careful when moving over slopes. • Do not park the stand on slopes. • At the new location, press the brake to secure the stand in position. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 69 Operation Transportposition: Moving position: Position de transport: Posición de transporte: Transportrichtung Transport direction Direction de transport Dirección de transporte G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 70: Checklist

    Warning! If a function fails, you must not use this instrument for safety reasons. Cor- rect the fault (see the "Troubleshooting table“) or contact our service dept. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 71 Operation G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 72: Using The Display And Key Field

    If you have selected any of the modes and do not press a key in the key field, the program will return to the basic mode after 20 seconds. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 73 Press the "MANUAL" key once again to reactivate electronic control; the display in the display and key field then returns to the basic mode. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 74 Make sure never to change the settings of another user. It is therefore ad- visable that you only use your own user ID for your work. Remember that all settings made are stored under the user ID currently selected. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 75 You can continue to use the illumination system. The lamp brightness, however, is automatically set to a fixed value. You can switch the illumi- nation on and off on the foot control panel. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 76: Operating The Opmi® Vario

    If no key is pressed within 20 seconds, the program auto- matically returns to the basic mode. Note: The illustration shows the factory-adjusted default values to which you can reset your specific settings if required. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 77 To set the instrument to these stored focusing and zoom values, press keys C and D on the foot control panel or the freely configurable keys on the handgrips, see configuration mode 1. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 78 Saving parameter settings As soon as you have entered a parameter setting, it is saved under the current user ID. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 79 Be extremely careful when changing these settings. You should change settings only under your own user ID. Notify all users of any changes, or make sure that each user only works under his own user ID. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 80 For further details, please see configura- tion mode 2. Saving parameter settings As soon as you have entered a parameter setting, it is saved under the current user ID. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 81 Path: To return to the basic mode, press the "MODE" key in the speed mode or do not press any key for 20 seconds. To switch to configuration mode 1, simultaneously press the "MODE" and "STORE" keys. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 82 At each press of the "∇" key, you advance in clockwise direction. At each press of the "∆" key, you advance in counterclockwise direction: → FOC-MEM → ZOOM-MEM → ↑ ↓ ← XY-RES ← AUX ← PHOTO ← G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 83 Path: To return to the basic mode, press the "MODE" key in configuration mode 1 or do not press any key for 20 seconds. To switch to configuration mode 2, simultaneously press the "MODE" and "STORE" keys. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 84 If you press key (1) for less than 2 seconds, a focus positioning run is started; if you press this key longer than 2 seconds, the current focus po- G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 85 Path: To return to the basic mode, press the "MODE" key in configuration mode 1 or do not press any key for 20 seconds. To switch to configuration mode 2, simultaneously press the "MODE" and "STORE" keys. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 86 In configuration mode 2, the top section (SPEED FUNCT:) of the display and key field permits you to select a high, medium or low value for the dy- namic change of speed, or to deselect dynamic speed control: HIGH MEDIUM G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 87 Path: To return to the basic mode, press the "MODE" key in the configu- ration mode or do not press any key for 20 seconds. To access the next configuration mode 3, simultaneously press the "MODE" and "STORE" keys. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 88 If you have selected a relatively high basic brightness in the basic mode, the automatic function may not be able to sufficiently compensate for re- duced transmission of the optical system. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 89 Path: To return to the basic mode, press the "MODE" key in the configu- ration mode or do not press any key for 20 seconds. To access the next configuration mode 3, simultaneously press the "MODE" and "STORE" keys. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 90 Be extremely careful when changing these settings. You should change settings only under your own user ID. Notify all users of any changes, or make sure that each user only works under his own user ID. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 91 Path: To return to the basic mode, press the "MODE" key in the configu- ration mode or do not press any key for 20 seconds. To access the next configuration mode 3, simultaneously press the "MODE" and "STORE" keys. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 92 Be extremely careful when changing these settings. You should change settings only under your own user ID. Notify all users of any changes, or make sure that each user only works under his own user ID. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 93 Path: To return to the basic mode, press the "MODE" key in the configu- ration mode or do not press any key for 20 seconds. To access configuration mode 1 from configuration mode 3, simul- taneously press the "MODE" and "STORE" keys. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 94 Be extremely careful when changing these settings. You should change settings only under your own user ID. Notify all users of any changes, or make sure that each user only works under his own user ID. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 95 Path: To return to the basic mode, press the "MODE" key in the configu- ration mode or do not press any key for 20 seconds. To access configuration mode 1 from configuration mode 3, simul- taneously press the "MODE" and "STORE" keys. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 96 If no key is pressed within 20 seconds, the program auto- matically returns to the basic mode. Note: The illustration shows the factory-adjusted default values to which you can reset your specific settings if required. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 97 Be extremely careful when changing these settings. You should change settings only under your own user ID. Notify all users of any changes, or make sure that each user only works under his own user ID. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 98 – during adjustment of the lamp brightness, when brightness level 1.0 has been reached. Saving parameter settings As soon as you have entered a parameter setting, it is saved under the current user ID. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 99 For further details, please see configura- tion mode 1. Saving parameter settings As soon as you have entered a parameter setting, it is saved under the current user ID. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 100 Path: To return to the basic mode, press the "MODE" key in the speed mode or do not press any key for 20 seconds. To switch to configuration mode 1, simultaneously press the "MODE" and "STORE" keys. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 101 At each press of the "∇" key, you advance in clockwise direction. At each press of the "∆" key, you advance in counterclockwise direction: → XYZ-RES → FOC-MEM → XY-INV → ↑ ↓ ← XY-RES ← AUX ← PHOTO ← SDI ← G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 102 You can stop this process at any point by briefly tipping on the appropriate key on the foot control panel (C or D) or on one of the direction keys (joystick or one of the two rocker switches). G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 103 Operation Triggers an AUX signal, e.g. for switching an external unit on or off. PHOTO Triggers the shutter of a 35 mm camera, if connected. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 104 In configuration mode 1, the bottom section (SPEED FUNCT:) of the dis- play and key field permits you to select a high, medium or low value for the dynamic change of speed, or to deselect dynamic speed control: HIGH MEDIUM G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 105 Path: To return to the basic mode, press the "MODE" key in configuration mode 1 or do not press any key for 20 seconds. To switch to configuration mode 2, simultaneously press the "MODE" and "STORE" keys. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 106 Be extremely careful when changing these settings. You should change settings only under your own user ID. Notify all users of any changes, or make sure that each user only works under his own user ID. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 107 Path: To return to the basic mode, press the "MODE" key in the configu- ration mode or do not press any key for 20 seconds. To access the next configuration mode 3, simultaneously press the "MODE" and "STORE" keys. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 108: Operating The Opmi® Pro Magis

    If no key is pressed within 20 seconds, the program auto- matically returns to the basic mode. Note: The illustration shows the factory-adjusted default values to which you can reset your specific settings if required. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 109 Be extremely careful when changing these settings. You should change settings only under your own user ID. Notify all users of any changes, or make sure that each user only works under his own user ID. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 110 Press the "MODE" key to access the speed mode. SPEED: Changing the settings The speed mode is the user interface where you can select the speeds of the surgical microscope functions. STORE MANUAL MODE G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 111 Path: To return to the basic mode, press the "MODE" key in the speed mode or do not press any key for 20 seconds. To switch to configuration mode 1, simultaneously press the "MODE" and "STORE" keys. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 112 As soon as you have entered a parameter setting, it is saved under the current user ID. Path: To return to the basic mode, press the "MODE" key in configuration mode 1 or do not press any key for 20 seconds. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 113 Operation G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 114: What To Do In An Emergency

    • Switch the ceiling mount (floor stand) back on. Adjust the brightness of the illumination on the display field (6) as required. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 115 Operation G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 116: Failure Of The Xenon Lamp

    (backup lamp) into the beam path. • Push the lamp module all the way back into the lamp housing. • Turn the stand back on at the power switch. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 117 Operation G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 118: Failure Of Lamp Control

    If the magnetic clutches fail (magnetic clutches are locked), you can man- ually position the articulated arm including the microscope by overcoming the locking effect of the magnetic clutches. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 119 Operation G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 120 Operation G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 121: Maintenance / Further Information

    Trouble-shooting table Changing the halogen lamp Changing the xenon lamp module Care of the unit Sterilization Disinfecting the control keys Ordering data Accessories Spare parts Technical data Ambient requirements G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 122: Trouble-Shooting Table

    OR illuminator. Contact service dept. Yellow indicator lamp in dis- Main lamp and backup lamp Change lamp or use backup play field blinks. defective. lamp module. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 123 Stand wobbles. Floor not level. Stand base Turn base slightly. The artic- not ideally positioned. ulated arm should be at right angles to the unstable axis. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 124 (floor stand) or foot control panel. Halogen lamp not properly Press halogen lamp into page 126 plugged into lamp socket. lamp socket. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 125 Failure of stand electronics. Illuminate surgical field using an OR illuminator. Contact service dept. ® OPMI Vario on S8 ceiling mount (floor stand) Problem Possible cause Remedy Incorrect display of total Eyepiece magnification has Enter current eyepiece magnification.
  • Page 126: Changing The Halogen Lamp

    Press button (2), the lamp module is slightly ejected. Remove the lamp module (1). • Remove the defective halogen lamp from its spring-loaded mount. • Pull the ceramic base (8) from the contact pins of the halogen lamp. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 127 Maintenance / Further information G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 128 If the main lamp (4) has failed, remove it and replace it by the backup lamp (5). Install the new hal- ogen lamp instead of the backup lamp. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 129 Maintenance / Further information G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 130: Changing The Xenon Lamp Module

    Pack the old lamp module (5) in the transport package of the new lamp module. Fill in the enclosed return card and send the old lamp module to the nearest Zeiss service agency. Note: •...
  • Page 131 Maintenance / Further information G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 132: Care Of The Unit

    Wipe off any residue with a mixture of 50% ethyl alcohol and 50% distilled water plus a dash of household dish-washing liquid. Sterilization The asepsis sets available from Carl Zeiss contain rubber caps, sleeves and grips which can be sterilized in autoclaves. We recommend the follo- wing program for sterilization: Sterilization temperature: 134°...
  • Page 133: Disinfecting The Control Keys

    Disinfecting the control keys To be able to use the unit in ORs, we recommend disinfecting it using MELISEPTOL disinfectant solution (B. Braun, Melsungen AG). Carl Zeiss can supply MELISEPTOL and you can obtain it in many countries from B. Braun representatives.
  • Page 134: Ordering Data

    Please observe the following: Only operate the instrument with the accessories included in the delivery package. If you want to use other accessories, make sure that Carl Zeiss or the manufacturer of the accessories has proved and confirmed that these accessories meet the respective technical safety standards and can be used without risk.
  • Page 135: Spare Parts

    Backup lamp module, complete with 2 xenon lamps in 1090-805 transport case and return card; in exchange for a re- turned module with defective xenon lamps Xenon lamp module, complete with 2 xenon lamps 1087-810 (new item) G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 136: Technical Data

    Maintenance / Further information Technical data S8 floor stand Mechanical data Suspension arm Length...850 mm Swivel angle...270° Vertical lift...± 360 mm Carrier arm Length...450 mm Swivel angle...360° Stand height 1880 mm Base (dimensions) 850 x 790 mm Admissible max. load on...
  • Page 137 IEC 601-1/EN 60 601-1/ UL 2601- standard 1; CSA C22.2 No 601.1 Protection class I, type B equipment complying with EN 60601-1-2: 1993 requirements Class A (hospital) G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 138 Color temperature: approx. 5000 K Rated power: 180 W, rated voltage approx. 14.5 V, rated current: 12 A, ignition voltage: 20 kV Backup lamp in lamp housing, manual activation. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 139: Ambient Requirements

    500 hPa...1,060 hPa The unit meets the essential requirements stipulated in Annex I to the 93/42/EEC Directive governing medical de- vices. The unit is marked with: Subject to change. G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 140 Maintenance / Further information G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 141 Control functions for OPMI® PRO magis Control functions for OPMI® Vario Control functions for OPMI® VISU 150 / 200 Control keys, disinfecting Description Description of the modules Directive 93/42/EEC G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 142 Inverting the SDI movement Inverting the X-Y coupling movement ISO standard, future Key "B" Key to symbols Keys Keys on the right of the display field 18, 42 KK 40 filter G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 143 Operating VISU 150 and VISU 200 Operation Operation, requirements Ordering data PHOTO 82, 84, 101, 112 Photo Position, saving of Positioning run Positioning the stand Power outlet Power switch G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 144 Re-equipping the instrument Release bar Relocating the stand Repeat function Requirements for operation Risk of fire Rotary and tilt movements Row of keys, operation S8 floor stand 9, 10 Safety 12, 122 Safety check Safety devices Safety standards Securing and adjustment screws...
  • Page 145 XY movement 82, 84 XY recentering XY-INV XY-RES XYZ-RES 44, 48 Yellow indicator lamp Yellow LED for "MANUAL" Z movement 82, 84 Zoom memory 80, 99, 110 Zoom speed G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 146 Index G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 147 G-30-1346-en S8 Floor Stand Issue 4.0 Printed on 05. 06. 2001...
  • Page 148 Carl Zeiss Geschäftsbereich Chirurgische Geräte D-73446 Oberkochen Fax: 07364-204455 G-30-1346-en Printed in Germany MS-TD...

Table of Contents