automatic power-on self-test with error message

Synchronization

with ECG signal of either polarity via monitor:
indicated by yellow LED
trigger marks on monitor display
in synchronized mode delivery of defibrillation pulse delayed by 15 ms max. after trigger pulse
from monitor (delay between R-wave and sync pulse: refer to operator's manual of patient
monitor)

Pacemaker

operating modes: Demand, Fix
pacer rate: 30 to 200 ppm, adjustable in steps of 5 ppm, ± 1%
pacer output: 0 to 200 mA ± 10% (0 to 500 Ohms), adjustable in steps of 5 mA
pulse width: 20 ms, ± 0,5%
pulse shape: monophase square-wave pulse
recovery time of 50 ms after pulse delivery

Power Supply

from exchangeable NiCd battery
rated voltage 12 V
rated capacity 2.0 Ah
battery charges either in the device or in the external charging unit ASU 3000 When charged in the
device, power supply from connected Dash
charging voltage 9 to 18 V, DC voltage
-
charging current 320 mA to 490 mA
-
max. charge time of a depleted battery 6 hours
-
operating time per battery charge approx. 60 defibrillation shocks with 360 Joules each (into
-
50 Ohms)

Operational Readiness

seconds, incl. automatic self-test

Operating Position

horizontal

Type of Protection

protected against penetration of spray water and small objects: IP 23 (battery inserted, cable connected)

Environment

Operation

operation under the following conditions considered to be normal:
Page 52 of 72
2000, Dash
Servicing Manual Dash Responder V1.0
3000, or Dash
4000 patient monitor:
Revision F
2004768-001