GE Medical Systems
Information Technologies
•
automatic power-on self-test with error message
Synchronization
with ECG signal of either polarity via monitor:
•
indicated by yellow LED
•
trigger marks on monitor display
•
in synchronized mode delivery of defibrillation pulse delayed by 15 ms max. after trigger pulse
from monitor (delay between R-wave and sync pulse: refer to operator's manual of patient
monitor)
Pacemaker
•
operating modes: Demand, Fix
•
pacer rate: 30 to 200 ppm, adjustable in steps of 5 ppm, ± 1%
•
pacer output: 0 to 200 mA ± 10% (0 to 500 Ohms), adjustable in steps of 5 mA
•
pulse width: 20 ms, ± 0,5%
•
pulse shape: monophase square-wave pulse
•
recovery time of 50 ms after pulse delivery
Power Supply
from exchangeable NiCd battery
•
rated voltage 12 V
•
rated capacity 2.0 Ah
•
battery charges either in the device or in the external charging unit ASU 3000 When charged in the
device, power supply from connected Dash
charging voltage 9 to 18 V, DC voltage
-
charging current 320 mA to 490 mA
-
max. charge time of a depleted battery 6 hours
-
operating time per battery charge approx. 60 defibrillation shocks with 360 Joules each (into
-
50 Ohms)
Operational Readiness
•
seconds, incl. automatic self-test
Operating Position
horizontal
Type of Protection
protected against penetration of spray water and small objects: IP 23 (battery inserted, cable connected)
Environment
Operation
•
operation under the following conditions considered to be normal:
Page 52 of 72
2000, Dash
Servicing Manual Dash Responder V1.0
3000, or Dash
4000 patient monitor:
Revision F
2004768-001