Fluoride, Pipet Method; Measuring Hints - Hach POCKET COLORIMETER II Instruction Manual

Analysis system
Hide thumbs Also See for POCKET COLORIMETER II:
Table of Contents

Advertisement

Fluoride, Pipet Method

For water, wastewater, and seawater
SPADNS Method* USEPA Accepted (distillation required)

Measuring Hints

Remove liquid and fingerprints from the sample cells with a soft, dry cloth
before placing in the instrument.
If samples cannot be analyzed immediately, see
page
1—30.
SPADNS Reagent contains sodium arsenite. Final solutions will contain
arsenic (D004) in sufficient concentration to be regulated as a hazardous waste
for Federal RCRA.
Note:
The Pocket Colorimeter II is designed to measure solutions contained in sample cells.
dip the meter in the sample or pour the sample directly into the cell holder.
DO NOT
* Adapted from Standard Methods for Examination of Water and Wastewater.
(0.02 to 2.00 mg/L F
Sampling and Storage on
1—17
)
Method 8029