Page 3
The reader should consult National Instruments if errors are suspected. In no event shall National Instruments be liable for any damages arising out of or related to this document or the information contained in it.
Page 4
Conventions The following conventions are used in this manual: <> Angle brackets that contain numbers separated by an ellipsis represent a range of values associated with a bit or signal name—for example, DIO<3..0>. » The » symbol leads you through nested menu items and dialog box options to a final action.
Installing the SCXI-1581 ....................1-2 Unpacking......................1-2 Installing the SCXI-1581 Module into the SCXI Chassis.......1-3 Connecting the SCXI-1581 to an E Series DAQ Device for Multiplexed Scanning in an SCXI Chassis..........1-4 Connecting the SCXI-1581 to an E Series DAQ Device for Multiplexed Scanning in a PXI Combination Chassis......1-6 Configuring and Self-Testing ..................1-7...
Page 6
RTD Measurement Errors ................B-6 Thermistors........................B-7 Resistance/Temperature Characteristic of Thermistors ........B-9 Thermistor Measurement Circuits ..............B-10 Connecting Resistive Devices to the SCXI-1581............B-11 4-Wire Configuration ..................B-12 2-Wire Configuration ..................B-13 3-Wire Resistive Sensor Connected in a 2-Wire Configuration ..... B-14 Lead-Resistance Compensation Using a 3-Wire Resistive Sensor and Two Matched Current Sources..............
You can use the SCXI-1581 in any application that requires 100 µA fixed current excitation. The SCXI-1581 is configured using Measurement & Automation Explorer (MAX). What You Need to Get Started To set up and use the SCXI-1581, you need the following items: ❑ SCXI-1581 and documentation ❑...
CD. Installing the SCXI-1581 The following section describes how to install the SCXI-1581 for use with SCXI chassis and E Series DAQ devices. Caution Refer to the Read Me First: Radio-Frequency Interference document before removing equipment covers or connecting or disconnecting any signal wires.
Power off the SCXI chassis. Do not insert the SCXI-1581 module into a chassis that is powered on. Insert the SCXI-1581 into an open slot in the SCXI chassis. Refer to the SCXI Quick Start Guide for information on the correct order to install modules.
3 SCXI Module Thumbscrews 2 SCXI Chassis Slots Figure 1-1. Installing the SCXI-1581 Module To complete the SCXI-1581 installation, follow the steps in the appropriate sections that follow. Connecting the SCXI-1581 to an E Series DAQ Device for Multiplexed Scanning in an SCXI Chassis...
Page 13
Consult the SCXI chassis documentation, other SCXI module documentation, and E Series DAQ device documentation for additional instructions and cautions. You should have already installed the SCXI-1581 module and any other SCXI modules in the chassis according to their installation instructions.
Configuring and Self-Testing section to configure the SCXI-1581 module(s) for multiplexed mode operation. Connecting the SCXI-1581 to an E Series DAQ Device for Multiplexed Scanning in a PXI Combination Chassis In this configuration, you can multiplex the input channels of the SCXI analog-input modules into a single channel of the E Series DAQ device in the combination PXI chassis.
NI-DAQmx Devices. Configuring SCXI Using Traditional NI-DAQ Use MAX to configure and test the SCXI-1581. If you need help during the configuration process, open the Measurement & Automation Help by selecting Help Topics from the Help menu. Complete the following steps to configure the SCXI system.
Page 16
NI-DAQ section. If you selected auto-detection, select the communication path and click Next. Select the cabled module by clicking the module in the list box and clicking Finish. You have completed configuring the SCXI-1581 using Traditional NI-DAQ. SCXI-1581 User Manual ni.com...
Chapter 1 Installing and Configuring the SCXI-1581 Manually Adding and Configuring Modules in Traditional NI-DAQ If you did not auto-detect the SCXI modules, you must manually add and configure each of the modules. Complete the following steps to manually add modules.
Self-Test Verification section for troubleshooting steps. Configuring SCXI Using NI-DAQmx Use MAX to configure and test the SCXI-1581. If you need help during the configuration process, open the Measurement & Automation Help by selecting Help Topics from the Help menu. Complete the following steps to configure the SCXI system.
Chapter 1 Installing and Configuring the SCXI-1581 Manually Adding and Configuring Modules in NI-DAQmx You must add and configure each of the modules in the SCXI Chassis Configuration window. If you are not in the SCXI Chassis Configuration window, right-click on the chassis in which the module is installed and select Properties.
NI-DAQmx section and make sure the correct module is in the specified slot. Delete the incorrect module as described in the Removing the SCXI-1581 from Measurement & Automation Explorer section. Add the correct module as described in the Manually Adding and Configuring Modules in NI-DAQmx section.
MAX: Display the list of installed devices and interfaces by clicking the + next to the Devices and Interfaces icon. If the SCXI-1581 is configured under NI-DAQmx, remove the module under NI-DAQmx Devices. If the SCXI-1581 is configured under Traditional NI-DAQ, remove the module under Traditional NI-DAQ Devices.
SCXI-1581 module from an SCXI chassis, complete the following steps while referring to Figure 1-3: Power off the SCXI chassis. Do not remove the SCXI-1581 module from a chassis that is powered on. If the SCXI-1581 is the cabled module, disconnect the cable running from the SCXI-1581 to the E Series DAQ device.
3 SCXI Module Thumbscrews 6 SCXI Chassis Figure 1-3. Removing the SCXI-1581 Removing the SCXI-1581 from Measurement & Automation Explorer To remove a module from MAX, complete the following steps after launching MAX: Display the list of installed devices and interfaces by expanding Devices and Interfaces.
Page 24
Chapter 1 Installing and Configuring the SCXI-1581 Right-click the module or chassis you want to delete and click Delete. Note Deleting the SCXI chassis deletes all modules in the chassis. All configuration information for these modules is also lost. You are presented with a confirmation window. Click Yes to continue deleting the module or chassis or No to cancel this action.
The current sources on the SCXI-1581 continuously provide 32 channels of 100 µA current excitation. These current sources are on whenever the SCXI chassis is powered-on. The current sources on the SCXI-1581 are designed to be accurate to within ±0.05% of the specified value with a temperature drift of no more than ±5 ppm/°C.
Chapter 2 Using the SCXI-1581 Table 2-1. Front Signal Pin Assignments Front Connector Diagram Pin Number Column A Column B Column C EX0– EX0+ EX1– EX1+ Column EX2– EX2+ EX3– EX3+ RSVD EX4– EX4+ RSVD EX5– EX5+ RSVD EX6– EX6+ RSVD EX7–...
The rear signal connector, shown in Figure 2-1, is used for analog signal connectivity and communication between the SCXI-1581 and the E Series DAQ device. Grounding signals AIGND and OUTREF provide reference signals needed in the various analog input referencing modes on the E Series DAQ device.
SCXI modules to the E Series DAQ device. The SCXI-1581 routes multiplexed signals from other SCXI modules to the E Series DAQ device even though the SCXI-1581 does not have any analog-input channels.
CH0 pins on its rear signal connector shown in Figure 2-1. Because the SCXI-1581 does not have any input channels to scan, multiplexed scanning is the only scanning mode that is possible when the SCXI-1581 is the cabled module. If the module you are scanning is directly cabled to the E Series DAQ device, the module routes its input signals through the CH0 pins on its rear signal connector.
Page 31
Specifications This appendix lists the specifications for the SCXI-1581 modules. These specifications are typical at 25 °C unless otherwise noted. Stability Recommended warm-up time ....10 minutes Excitation Channels..........32 single-ended outputs Current output ........100 µA Accuracy ..........±0.05% Temperature drift ........
Physical Dimensions ..........3.0 by 17.2 by 20.3 cm (1.2 by 6.9 by 8.0 in.) Safety The SCXI-1581 meets the requirements of the following standards for safety and electrical equipment for measurement, control, and laboratory use: • IEC 61010-1, EN 61010-1 •...
RTD at 0 °C. For example, a 100 Ω platinum RTD with α = 0.003911 has a resistance of 139.11 Ω at 100 °C. Figure B-1 displays a typical resistance-temperature curve for a 100 Ω platinum RTD. SCXI-1581 User Manual ni.com...
To create an RTD virtual channel using traditional NI-DAQ, complete the following steps: Launch MAX. Right-click Data Neighborhood and select Create New. Select Traditional NI-DAQ Virtual Channel and click Finish. Select Analog Input from the drop-down listbox and click Next. SCXI-1581 User Manual ni.com...
For a platinum RTD at 0 °C with α = 0.00385, the lead resistance equates to an error of approximately 0.6 W 1.6 °C ----------------------------- 0.385 W/°C The Connecting Resistive Devices to the SCXI-1581 section describes different ways of connecting resistive devices to the SCXI system. SCXI-1581 User Manual ni.com...
Page 42
Figure B-3. Resistance-Temperature Curve for a 2,252 Thermistor The thermistor has been used primarily for high-resolution measurements over limited temperature ranges. However, continuing improvements in thermistor stability, accuracy, and interchangeability have prompted increased use of thermistors in a variety of applications. SCXI-1581 User Manual ni.com...
Thermistors, with their small size and high resistance, are particularly prone to these self-heating errors. Manufacturers typically specify this self-heating as a dissipation constant, which is the power required to heat the thermistor 1 °C from ambient temperature (mW/°C). The dissipation SCXI-1581 User Manual B-10 ni.com...
You can use the SCXI-1300 terminal block to make signal connections to the SCXI-1581 and SCXI-1102 modules. When using the SCXI-1300 terminal block, terminals EX<0..31>+ and EX<0..31>– map to terminals CH<0..31>+ and CH<0..31>– respectively on the SCXI-1300 terminal block.
The main disadvantage of the 4-wire connection is the greater number of field wires required. SCXI-1581 SCXI-1300 EX0+ EX0– SCXI-1300 SCXI-1102 CH0+ – CH0– , and R are not required to be equal Figure B-5. 4-Wire Resistive Sensor Connected in a 4-Wire Configuration SCXI-1581 User Manual B-12 ni.com...
The resulting measurement error is: VE = R An advantage of this configuration is that it only requires a single jumper wire from the SCXI-1581 EX0+ terminal to the SCXI-1102B/C CH0+ terminal. SCXI-1300 SCXI-1581 EX0+ EX0–...
Measuring Temperature with Resistive Transducers The effectiveness of this method depends on the matching of the current sources. Each current source on the SCXI-1581 has an accuracy of ±0.05%. This accuracy results in a worst-case matching of ±0.1%. As an example of the error introduced by this mismatch, consider a 3-wire resistive sensor with lead resistances of 10 Ω.
Common Questions This appendix lists common questions related to the use of the SCXI-1581. Which version of NI-DAQ works with the SCXI-1581, and how do I get the most current version of NI-DAQ? NI recommends NI-DAQ 6.9.1 or later. With earlier versions of NI-DAQ, you cannot auto-detect or list the SCXI-1581 in the chassis as a configured module.
Page 54
If you require isolated current outputs, consider the SCXI-1124 module instead. Can I modify the SCXI-1581 circuitry to generate current at a level different than 100 µA? No. Do not attempt to modify any circuitry in the SCXI-1581.
Which digital lines are unavailable on the E Series DAQ device if I am cabled to an SCXI-1581 module? Table C-1 shows the digital lines that are used by the SCXI-1581 for communication and scanning. These lines are unavailable for general-purpose digital I/O if the SCXI-1581 is connected to the DAQ device.
Page 56
Does short-circuiting a current-output channel do any damage to the SCXI-1581? No. The SCXI-1581 delivers 100 µA into any load from 0 Ω to 100 kΩ. Does open-circuiting a current-output channel damage the SCXI-1581? What is the open-circuit voltage level? No.
Page 57
Isothermal Terminal Block Installation Guide for more information. As with the SCXI-1300 and BNC-2095, you cannot measure the CJC sensor. Can I configure the SCXI-1581 in MAX for use in a chassis controlled by an SCXI-1200? Yes. However, remember that you cannot auto-detect modules in a chassis controlled by an SCXI-1200.
Page 62
DAQ operation to begin FIFO first-in first-out memory buffer filtering a type of signal conditioning that allows you to remove unwanted frequency components from the signal you are trying to measure front signal connector full-scale range SCXI-1581 User Manual ni.com...
Page 64
(signal) NI-DAQ the driver software needed in order to use National Instruments E Series DAQ devices and SCXI components SCXI-1581 User Manual ni.com...
Page 66
SCANCLK scan clock signal used to increment to the next channel after each E Series DAQ device analog-to-digital conversion SCXI Signal Conditioning eXtensions for Instrumentation SCXI-1581 User Manual ni.com...