Technical Alarms (INOPs)

5
Technical alarms, also known as INOPs, indicate that the measuring device cannot measure reliably. If
an INOP interrupts monitoring (for example,
of the measurement numeric. An INOP tone sounds at the Cableless Measurement Device only when
there is no SRR connection to a host.
Some INOPs have a severity indication:
INOP.
NOTE
The IntelliVue Cableless Measurements do not provide physiological alarms locally on the
IntelliVue CL SpO

Display of INOPs

INOP messages are shown in black on a light gray background and coded according to their severity.
Standard INOPs do not have a symbol, red and yellow INOPs have exclamation marks beside the
alarm message:
An INOP message that appears is automatically highlighted. Use the  key to acknowledge the
message. An acknowledge message is displayed at the bottom of the screen and highlighted. Press 
again to confirm the acknowledgment of the INOP. After the confirmation, any ongoing INOP
message is displayed in the icon tray on the right side of the screen.

Acknowledging an INOP

WARNING
If the monitor is mechanically damaged, or if it is not working properly, do not use it for any
monitoring procedure on a patient. Contact your service personnel.
To acknowledge an INOP, select the INOP message and press the  key.
Pod and the IntelliVue CL NBP Pod.
2
!!!
!!
for red INOPs and
NBP Interrupted
), there will be a question mark in place
!!
for a medium severity INOP and
for yellow INOPs.
5
!!!
for a high severity
47