Process Ph Calibration - Hanna Instruments Groline HI981412 Instruction Manual

Ph dosing system
Table of Contents

Advertisement

• The "CAL" tag is displayed in measurement mode.
• If the buffer is not recognized, "---- WRONG" message is displayed. It is recommended to change
the solution and / or clean the electrode.
• Press CAL key to exit calibration.
Note: If high accuracy is required, a two-point calibration is recommended.

10.2. PROCESS pH CALIBRATION

Prior to performing a process calibration, determine the value using a reference meter. Make a note
of the value.
Note: To enter process calibration, the user must have the controller and probe calibrated
previously on the meter. The "CAL" tag must be on.
pH process calibration is a single point calibration performed while the probe remains installed in the
process. The value can be set ± 0.50 around measured pH.
• Press CAL key to enter calibration mode. When the first buffer value is displayed, press one of the
arrow keys to enter process calibration.
• Press the arrow keys again to adjust process calibration value to what was determined with the
hand held measurement. "PROCESS" message is scrolled on the bottom of the LCD screen.
33

Advertisement

Table of Contents
loading

Table of Contents