7. SETUP

• Short press MENU key to move to next item in menu.
• Long press MENU key to exit.
• Press arrow keys to change the values.
• Short press MENU key to automatically save modified values.
Table below presents an overview of the menu with ranges and factory set defaults.
Parameter
(Scrolled message)
Control
Control type
Control mode
Set point
Hysteresis (on/oFF only)
Band (Proportional only)
Start up delay (seconds)
Overtime alarm (minutes)
Flow rate (Liter/hour)
Flow rate (Gallon/hour)
Level alarm
High alarm
High alarm value (pH)
Low alarm
Low alarm value (pH)
Temperature unit
Flow rate unit
Language
*Available range changes based upon other settings.
High alarm must be set higher than low alarm. If low alarm was set 7 pH, then high alarm range is
7.1 to 14 pH.
Range / Option
Auto or oFF
on/oFF or Proportional
Hi or Lo
4.00 to 10.00
0.10 to 1.00 pH
0.10 to 2.00 pH
0-600
oFF, 0-180
0.5 to 3.5
0.13 to 0.92
diS or En
diS or En
0 to 14.00 pH*
diS or En
0 to 13.90 pH*
°C or °F
L.H or GAL.H
En (English), ES (Spanish),
Fr (French), Pt (Portuguese),
nL (Dutch), dE (German)
Default settings
Auto
on/oFF
Hi
7.20
0.50
1.0
60
30
1.0
0.26
diS
En
10.00 pH
diS
4.00 pH
°C
L.H
En (English)
21