Output:polarity; Output:shunt; Output:polarity <Select>,(@<Ch_List>); Output:polarity? (@<Channel>) - HP VXI 75000 C Series User's And Scpi Programming Manual

Algorithmic closed loop controller
Table of Contents

Advertisement

OUTPut:POLarity

Parameters
Comments
Usage
OUTPut:POLarity?
Parameters
Comments

OUTPut:SHUNt

Chapter 6
OUTPut:POLarity < select >,(@<ch_list>)
channels in <ch_list>.
Parameter
Parameter
Name
Type
select
discrete (string)
ch_list
string
If the channels specified do not support this function, an error will be
generated.
Related Commands: INPut:POLarity, OUTPut:POLarity?
*RST Condition: OUTP:POL NORM for all digital channels
OUTP:POL INV,(@144)
OUTPut:POLarity? (@< channel >)
in <channel>.
Parameter
Parameter
Name
Type
channel
string
Channel must specify a single channel
Returned Value: returns one of NORM or INV. The type is string.
OUTPut:SHUNt < enable >,(@< ch_list >)
on Strain Bridge Completion SCPs. This can be used for diagnostic purposes and
characterization of bridge response.
sets the polarity on digital output
Range of
Values
NORMal | INVerted
100 - 163
invert output logic sense on channel 44
returns the polarity on the digital output channel
Range of
Values
100 - 163
adds shunt resistance to one leg of bridge
HP E1415 Command Reference
OUTPut
Default
Units
none
none
Default
Units
none
223

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vxi e1415a

Table of Contents