(Set Calibrated Gravity Constant)
This setting (e.g.,
where the scale was calibrated.
With the display showing
value displayed is acceptable, press the
Otherwise, press the
and proceed to the next prompt.
Allowable values are 9.700 through 9.900.
(SET OPERATED GRAVITY CONSTANT)
This setting (e.g.,
where the scale will be operated.
With the display showing
value displayed is acceptable, press the
Otherwise, press the
and proceed to the next prompt.
Allowable values are 9.700 through 9.900.
(Calibrate Scale)
Step 1. The display will change to show
Step 2. To skip calibration and save the previous settings, press the
calibration mode. The scale will turn off automatically.
Step 3. Otherwise, press the
can still select
Step 4. Press the
Step 5. Make certain the scale platform is empty and free of debris.
Step 6. The display will change to show the internal count.
Step 7. Press the
Step 8. The display will change to show the first calibration point
Step 9. While safely applying motion to the scale platform, load the first set of test weights
(equal to
on the scale, stop applying motion to the scale.
Step 10. Once motion has stopped and the weight is stable (STA annunciator will turn on),
the first weight reading will be saved, and the display will change to show the
second calibration point
Step 11. Again, while safely applying motion to the scale platform, load the remaining test
weights (add the remaining
test weights are on the scale, stop applying motion to the scale.
Step 12. Once motion has stopped and the weight is stable (STA annunciator will turn on),
the second weight reading will be saved, the display will change to show
and the scale will automatically turn off.
26
CALIBRATION, CONT.
) is the acceleration of gravity constant value for the location
, press the
or
keys to adjust the setting then press the
) is the acceleration of gravity constant value for the location
, press the
or
keys to adjust the setting then press the
or
keys to select
(n = exit calibration), skip calibration, and turn the scale off.
key to confirm.
key to save the zero-reference point.
kg or
lb) on the scale platform. When all the test weights are
0044-0391-0M Rev F solo Series Owner's Manual
key to show the current setting. If the
key and proceed to the next prompt.
key to show the current setting. If the
key and proceed to the next prompt.
.
(Y = start calibration). Note that you
kg or
lb.
kg or
lb) on the scale platform. When all the
key to save it
key to save it
key to exit the
kg or
lb.
,

Advertisement

Table of Contents
loading

This manual is also suitable for:

Solo-ri pd100Solo

Table of Contents