Pneumatic flow path
Pressure accuracy
Maximum static pressure variation at 10 cm H
ISO 80601-2-70:2015:
Tested with full face mask: ±0.5 cm H
Maximum dynamic pressure variation according to ISO 80601-2-70:2015
Machine with full face mask
Pressure
[cm H
O (hPa)]
2
4
8
12
16
20
0 (0.5 hPa)
2
10 BPM
0.5
0.5
0.5
0.5
0.5
1. Flow sensor
2. Blower
3. Pressure sensor
4. AAV (F20 connector only)
5. Vent
6. HumidX (N20, N30 and P10 connectors
only)
7. Mask
8. Air tubing
9. Machine
10. Inlet filter
O (10 hPa) according to
2
15 BPM
0.7
0.7
0.7
0.7
0.7
20 BPM
1.0
1.0
1.0
1.0
1.0